Steady-state pattern electroretinogram and short-duration transient visual evoked potentials in glaucomatous and healthy eyes.
Dilru C Amarasekera, Arthur F Resende, Michael Waisbourd, Sanjeev Puri, Marlene R Moster, Lisa A Hark, L Jay Katz, Scott J Fudemberg, Anand V Mantravadi
Summary
Steady-state pattern electroretinogram was effectively able to discern between glaucomatous and healthy eyes. Steady-state pattern electroretinogram may thus have a role as a clinically useful electrophysiological diagnostic tool.
Abstract
IMPORTANCE
This study evaluates two rapid electrophysiological glaucoma diagnostic tests that may add a functional perspective to glaucoma diagnosis.
BACKGROUND
This study aimed to determine the ability of two office-based electrophysiological diagnostic tests, steady-state pattern electroretinogram and short-duration transient visual evoked potentials, to discern between glaucomatous and healthy eyes.
DESIGN
This is a cross-sectional study in a hospital setting.
PARTICIPANTS
Forty-one patients with glaucoma and 41 healthy volunteers participated in the study.
METHODS
Steady-state pattern electroretinogram and short-duration transient visual evoked potential testing was conducted in glaucomatous and healthy eyes. A 64-bar-size stimulus with both a low-contrast and high-contrast setting was used to compare steady-state pattern electroretinogram parameters in both groups. A low-contrast and high-contrast checkerboard stimulus was used to measure short-duration transient visual evoked potential parameters in both groups.
MAIN OUTCOME MEASURES
Steady-state pattern electroretinogram parameters compared were MagnitudeD, MagnitudeD/Magnitude ratio, and the signal-to-noise ratio. Short-duration transient visual evoked potential parameters compared were amplitude and latency.
RESULTS
MagnitudeD was significantly lower in glaucoma patients when using a low-contrast (P = 0.001) and high-contrast (P < 0.001) 64-bar-size steady-state pattern electroretinogram stimulus. MagnitudeD/Magnitude ratio and SNR were significantly lower in the glaucoma group when using a high-contrast 64-bar-size stimulus (P < 0.001 and P = 0.010, respectively). Short-duration transient visual evoked potential amplitude and latency were not significantly different between the two groups.
CONCLUSIONS AND RELEVANCE
Steady-state pattern electroretinogram was effectively able to discern between glaucomatous and healthy eyes. Steady-state pattern electroretinogram may thus have a role as a clinically useful electrophysiological diagnostic tool.
Keywords
More by Dilru C Amarasekera
View full profile →Top Research in Diagnosis & Screening
Browse all →Efficacy of a Deep Learning System for Detecting Glaucomatous Optic Neuropathy Based on Color Fundus Photographs.
Dry eye disease and oxidative stress.
Central Corneal Thickness in the Ocular Hypertension Treatment Study (OHTS).
Discussion
Comments and discussion will appear here in a future update.