Efficacy, Safety, and Risk Factors for Failure of Standalone Ab Interno Gelatin Microstent Implantation versus Standalone Trabeculectomy.
Matthew B Schlenker, Husayn Gulamhusein, Ina Conrad-Hengerer, Alix Somers, Markus Lenzhofer, Ingeborg Stalmans, Herbert Reitsamer, Fritz H Hengerer, Iqbal Ike K Ahmed
Summary
There was no detectable difference in risk of failure and safety profiles between standalone ab interno microstent with MMC and trabeculectomy with MMC.
Abstract
PURPOSE
To compare the efficacy, safety, and risk factors for failure of standalone ab interno gelatin microstent implantation with mitomycin C (MMC) versus trabeculectomy with MMC.
DESIGN
International, multicenter, retrospective interventional cohort study.
PARTICIPANTS
Three hundred fifty-four eyes of 293 patients (185 microstent and 169 trabeculectomy) with no prior incisional surgery.
METHODS
Consecutive eyes with uncontrolled glaucoma underwent microstent or trabeculectomy surgery from January 1, 2011 through July 31, 2015 at 4 academic ophthalmology centers: Toronto, Canada; Frankfurt, Germany; Salzburg, Austria; and Leuven, Belgium.
MAIN OUTCOME MEASURES
Primary outcome measure was hazard ratio (HR) of failure, with failure defined as 2 consecutive intraocular pressure (IOP) readings of 17 mmHg without glaucoma medications (complete success) at least 1 month after surgery despite in-clinic interventions (including needling). Secondary outcome measures included IOP thresholds of 6 to 14 mmHg and 6 to 21 mmHg and same thresholds allowing for medications (qualified success), interventions, complications, and reoperations.
RESULTS
Baseline characteristics were similar, except more men (56% vs. 43%), younger patients (average, by 3 years), better preoperative visual acuity (22% vs. 32% with 0.4 logarithm of the minimum angle of resolution vision or worse), and more trabeculoplasty (52% vs. 30%) among microstent eyes. The adjusted HR of failure of the microstent relative to trabeculectomy was 1.2 (95% confidence interval [CI], 0.7-2.0) for complete success and 1.3 (95% CI, 0.6-2.8) for qualified success, and similar for other outcomes. Time to 25% failure was 11.2 months (95% CI, 6.9-16.1 months) and 10.6 months (95% CI, 6.8-16.2 months) for complete success and 30.3 months (95% CI, 19.0-∞ months) and 33.3 months (95% CI, 25.7-46.2 months) for qualified success. Overall, white ethnicity was associated with decreased risk of failure (adjusted HR, 0.49; 95% CI, 0.25-0.96), and diabetes was associated with increased risk of failure (adjusted HR, 4.21; 95% CI, 2.10-8.45). There were 117 and 165 distinct interventions: 43% and 31% underwent needling, respectively, and 50% of trabeculectomy eyes underwent laser suture lysis. There were 22 and 30 distinct complications, although most were transient. Ten percent and 5% underwent reoperation (P = 0.11).
CONCLUSIONS
There was no detectable difference in risk of failure and safety profiles between standalone ab interno microstent with MMC and trabeculectomy with MMC.
More by Matthew B Schlenker
View full profile →Intermediate Outcomes of a Novel Standalone Ab Externo SIBS Microshunt With Mitomycin C.
Artificial Intelligence Algorithms to Diagnose Glaucoma and Detect Glaucoma Progression: Translation to Clinical Practice.
Ab Externo SIBS Microshunt with Mitomycin C for Open-Angle Glaucoma: Three-Year Results as a Primary Surgical Intervention.
Top Research in Glaucoma Surgery
Browse all →Treatment Outcomes in the Primary Tube Versus Trabeculectomy Study after 1 Year of Follow-up.
Prospective, Randomized, Controlled Pivotal Trial of an Ab Interno Implanted Trabecular Micro-Bypass in Primary Open-Angle Glaucoma and Cataract: Two-Year Results.
A Schlemm Canal Microstent for Intraocular Pressure Reduction in Primary Open-Angle Glaucoma and Cataract: The HORIZON Study.