Three Dimensional Stimulus Source for Pattern Electroretinography in Mid- and Far-peripheral Retina.
Shresta Patangay, Zahra Derafshi, Thasarat S Vajaranant, Jason C Park, Elham Ghahari, J Jason McAnany, John R Hetling
Summary
Peripheral-field pERG responses are robust and repeatable; the unique response properties reflect differences between central and peripheral retina. Field-sector response ratios can be used to probe for sectoral dysfunction associated with disease.
Abstract
PURPOSE
The pattern electroretinogram (pERG) response reflects, in part, ganglion cell function. However, probing retinal ganglion cell (RGC) function in the mid- and far peripheral retina is difficult with conventional flat-panel pERG stimulus sources. A pattern stimulus source is presented for probing the peripheral retina. Peripheral pERG (ppERG) responses were evaluated versus luminance, reversal rate, and field subtended, and were compared with conventional pERG in healthy eyes.
METHODS
Eleven normally-sighted subjects were recruited. A hemispherical surface was used to present a reversing checkerboard pattern to the peripheral retina, from approximately 35° to 85° of visual field, in all directions. Responses to stimuli presented to peripheral field sectors (superior, nasal, inferior, temporal) were also recorded. Conventional pERG responses were recorded on the same day. Amplitudes and implicit times of waveform peaks were evaluated.
RESULTS
Robust pERG responses from peripheral retina resemble conventional pERG responses but with shorter implicit times and reduced positive component. Responses to high-luminance patterns include high-frequency components resembling flash ERG oscillatory potentials. Negative response component amplitudes increased with increasing pattern luminance, and decreased with increasing reversal rate.
CONCLUSIONS
Peripheral-field pERG responses are robust and repeatable; the unique response properties reflect differences between central and peripheral retina. Field-sector response ratios can be used to probe for sectoral dysfunction associated with disease.
TRANSLATIONAL RELEVANCE
The ppERG approach provides direct measurement of proximal retinal function beyond the fields probed by conventional perimetry and pERG, providing access to a relatively under studied part of the retina relevant to early stage glaucoma.
Keywords
Top Research in Visual Field
Browse all →Optical coherence tomography angiography: A comprehensive review of current methods and clinical applications.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Improving our understanding, and detection, of glaucomatous damage: An approach based upon optical coherence tomography (OCT).
Discussion
Comments and discussion will appear here in a future update.