Macular versus nerve fibre layer versus optic nerve head imaging for diagnosing glaucoma at different stages of the disease: Multicenter Italian Glaucoma Imaging Study.
Manuele Michelessi, Ivano Riva, Enrico Martini, Michele Figus, Paolo Frezzotti, Luca Agnifili, Gianluca Manni, Luciano Quaranta, Stefano Miglior, Chiara Posarelli, Stefano Fazio, Francesco Oddone
Summary
The diagnostic accuracy of MRW, pRNFL and macular analysis by Spectralis SD-OCT is overall good. MRW and pRNFL analysis performs statistically and clinically better than macular analysis to discriminate early glaucoma from healthy eyes.
Abstract
PURPOSE
To compare the diagnostic accuracy of minimum rim width (MRW), peripapillary retinal nerve fibre layer (pRNFL) and multilayered macular analysis by Spectralis SD-OCT (Heidelberg Engineering, Germany) in discriminating perimetric glaucoma at different stages of the disease from healthy eyes.
METHODS
In this multicentre, prospective, evaluation of diagnostic tests study, multilayered macular analysis and MRW and pRNFL were obtained from one eye of 197 glaucoma (76 early, 68 moderate and 53 advanced) and of 83 healthy controls from the Multicenter Italian Glaucoma Imaging Study (MIGIS). The reference standard for classifying eyes as glaucomatous and for staging the disease was the visual field. The main outcome measures were area under the ROC curve (AUC) and sensitivity at fixed specificity (95%).
RESULTS
Average MRW and average pRNFL showed the highest and similar diagnostic accuracy in both the whole study population (AUC 0.968 and 0.939) and early glaucoma (AUC 0.956 and 0.929). Among the macular parameters, the three innermost retinal layers combined as the Ganglion Cell Complex provided the highest diagnostic accuracy (AUC 0.931) in the whole population, which was statistically similar to that of average pRNFL but inferior to that of average MRW. Compared to both average MRW and pRNFL, all macular parameters showed statistically significant lower accuracy in early glaucoma, but accuracy in moderate and advanced glaucoma showed no statistically significant differences among three protocols.
CONCLUSION
The diagnostic accuracy of MRW, pRNFL and macular analysis by Spectralis SD-OCT is overall good. MRW and pRNFL analysis performs statistically and clinically better than macular analysis to discriminate early glaucoma from healthy eyes.
Keywords
More by Manuele Michelessi
View full profile →What Do We Really Know about the Effectiveness of Glaucoma Interventions?: An Overview of Systematic Reviews.
Diagnostic accuracy of optical coherence tomography for diagnosing glaucoma: secondary analyses of the GATE study.
The Glaucoma Italian Pediatric Study (GIPSy): The Long-term Effect of Topical Latanoprost on Central Corneal Thickness.
Top Research in Optic Nerve & Disc
Browse all →Efficacy of a Deep Learning System for Detecting Glaucomatous Optic Neuropathy Based on Color Fundus Photographs.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Inflammation in Glaucoma: From the back to the front of the eye, and beyond.
Discussion
Comments and discussion will appear here in a future update.