Semi-automated kinetic perimetry: Comparison of the Octopus 900 and Humphrey visual field analyzer 3 versus Goldmann perimetry.
Claudia Bevers, Gauthier Blanckaert, Keer Karel Van, Jean-François Fils, Evelien Vandewalle, Ingeborg Stalmans
Summary
Humphrey field analyzer 3 and Octopus 900 without RT vector are promising successors of Goldmann perimetry.
Abstract
PURPOSE
To evaluate the clinical usefulness and reproducibility of (semi-)automated kinetic perimetry of the Octopus 900 and Humphrey field analyzer 3 (HFA3) compared to Goldmann perimetry as reference technique.
METHODS
A prospective interventional study of two study groups, divided into three subgroups. The first study group consisted of 28 patients, performing one visual field examination on each of the three devices. A second group of 30 patients performed four examinations, one on Goldmann and three on Octopus 900 with the following testing strategies: (1) with reaction time (RT) vector, no headphone; (2) without RT vector, no headphone; and (3) without RT vector, with headphone. Comparisons for V4e and I4e stimuli were made of the mean isopter radius (MIR) and of the distances of the isopter to the central visual axis in four directions. Statistical analysis was made with the R software version 3.2.2.
RESULTS
For V4e stimuli, the mean isopter radius showed no statistic significant difference comparing Goldmann to HFA3 [p-value = 0.144; confidence interval (CI) -0.152 to 0.019] and comparing Goldmann to Octopus 900 without RT vector, either with (p-value = 0.347; CI -0.023 to 0.081) or without headphone (p-value = 0.130; CI -0.011 to 0.095). Octopus 900 with RT vector produced a significantly larger MIR for V4e stimuli in comparison to Goldmann (p-value < 0.001). I4e stimuli produced statistically significantly larger visual field areas when comparing HFA3 and Octopus 900 to Goldmann perimetry.
CONCLUSION
Humphrey field analyzer 3 and Octopus 900 without RT vector are promising successors of Goldmann perimetry.
Keywords
Top Research in Visual Field
Browse all →Optical coherence tomography angiography: A comprehensive review of current methods and clinical applications.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Improving our understanding, and detection, of glaucomatous damage: An approach based upon optical coherence tomography (OCT).
Discussion
Comments and discussion will appear here in a future update.