Improving Spatial Resolution and Test Times of Visual Field Testing Using ARREST.
Summary
Spending time trying to accurately measure visual field locations that have high variability is not productive.
Abstract
PURPOSE
Correctly classifying progression in moderate to advanced glaucoma is difficult. Pointwise visual field test-retest variability is high for sensitivities below approximately 20 dB; hence, reliably detecting progression requires many test repeats. We developed a testing approach that does not attempt to threshold accurately in areas with high variability, but instead expends presentations increasing spatial fidelity.
METHODS
Our visual field procedure Australian Reduced Range Extended Spatial Test (ARREST; a variant of the Bayesian procedure Zippy Estimation by Sequential Testing [ZEST]) applies the following approach: once a location has an estimated sensitivity of 17 dB to "defect", or (3) "defect" to blind. To test this approach we used an empirical database of progressing moderate-advanced 24-2 visual fields (121 eyes) that we "reverse engineered" to create visual field series that progressed from normal to the end observed field. ARREST and ZEST were run on these fields with test accuracy, presentation time, and ability to detect progression compared.
RESULTS
With specificity for detecting progression matched at 95%, ZEST and ARREST showed similar sensitivity for detecting progression. However, ARREST used approximately 25% to 40% fewer test presentations to achieve this result in advanced visual field damage. ARREST spatially defined the visual field deficit with greater precision than ZEST due to the addition of non-24-2 locations.
CONCLUSIONS
Spending time trying to accurately measure visual field locations that have high variability is not productive. Our simulations indicate that giving up attempting to quantify size III white-on-white sensitivities below 17 dB and using the presentations saved to test extra locations should better describe progression in moderate-to-advanced glaucoma in shorter time.
TRANSLATIONAL RELEVANCE
ARREST is a new visual field test algorithm that provides better spatial definition of visual field defects in faster test time than current procedures. This outcome is achieved by substituting inaccurate quantification of sensitivities <17 dB with new spatial locations.
Keywords
More by Andrew Turpin
View full profile →A Comparison between the Compass Fundus Perimeter and the Humphrey Field Analyzer.
Orientation of the Temporal Nerve Fiber Raphe in Healthy and in Glaucomatous Eyes.
The Proportion of Individuals Likely to Benefit from Customized Optic Nerve Head Structure-Function Mapping.
Top Research in Visual Field
Browse all →Optical coherence tomography angiography: A comprehensive review of current methods and clinical applications.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Improving our understanding, and detection, of glaucomatous damage: An approach based upon optical coherence tomography (OCT).
Discussion
Comments and discussion will appear here in a future update.