Determining Optic Nerve Cupping Using Optical Coherence Tomography (OCT) Versus a New Electronic Mobile Device.
Andrea G L Sarmento, Adriano Sarmento, Luiz L Souza, Luiz A Oliveira, Marcus F R Rios, Thiago A L Domingues, Jayter S Paula, Rodrigo P C Lira
Summary
We present a low-cost, reliable EMD that may serve as an alternative to subjectively measuring vertical CDR in face-to-face consultations and hence increase the general public's access to glaucoma screening.
Abstract
PURPOSE
We present a new device that may aid in glaucoma screening. The vertical cup-to-disc ratio (CDR) obtained from a new electronic mobile device (EMD) was compared with that obtained from a spectral-domain optical coherence tomography. In addition, results from examiner's evaluations were compared with the results mentioned above.
METHODS
A single-masked study was performed to evaluate the CDR results from 110 eyes from 110 subjects whose ages ranged between 18 and 60 years (39±13 y old), of whom 66 (60%) were female individuals. Vertical CDR was analyzed 5 times with each device in a single visit and evaluated by 3 different ophthalmologists. The Spearman correlation test and Bland-Altman analysis were used to evaluate the correlation and agreement between methods.
RESULTS
The CDR average using spectral-domain optical coherence tomography was 0.47±0.04, whereas the CDR average using EMD was 0.42±0.02. The clinical evaluation resulted in an average of 0.34±0.27. The Spearman test showed a strong correlation between vertical CDR obtained with both devices (r=0.8319; P0.001). Bland-Altman analysis showed good agreement between both devices as well as between EMD and the examiners (95% limits of agreement from -0.20 to 0.10 and from -0.15 to 0.30, respectively).
CONCLUSIONS
We present a low-cost, reliable EMD that may serve as an alternative to subjectively measuring vertical CDR in face-to-face consultations and hence increase the general public's access to glaucoma screening. This may be especially useful for patients who live in remote areas.
Top Research in Optic Nerve & Disc
Browse all →Efficacy of a Deep Learning System for Detecting Glaucomatous Optic Neuropathy Based on Color Fundus Photographs.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Inflammation in Glaucoma: From the back to the front of the eye, and beyond.
Discussion
Comments and discussion will appear here in a future update.