RETeval Portable Electroretinogram Parameters in Different Severity Stages of Glaucoma.
Yoshiyuki Kita, Gábor Holló, Tsunehiro Saito, Yousuke Momota, Ritsuko Kita, Kazushige Tsunoda, Akito Hirakata
Summary
The noninvasive, portable RETeval full-field ERG device may be useful to detect glaucoma in moderate-to-advanced stages.
Abstract
UNLABELLED
PRéCIS:: Four parameters of the noninvasive, portable RETeval electroretinogram (ERG) system were found to correlate with visual field mean deviation and optical coherence tomography (OCT) thickness parameters, and may therefore be suitable for glaucoma detection.
PURPOSE
To investigate the RETeval full-field ERG parameters for accuracy of separating glaucoma and normal eyes, and correlation with glaucoma severity.
PATIENTS AND METHODS
Sixty-two eyes of 62 primary open-angle glaucoma patients [visual field mean deviation (MD) range: -0.44 to -31.15 dB] and 39 eyes of 39 healthy controls underwent one RETeval test (photopic negative response protocol), OCT imaging, and Humphrey 30-2 visual field testing. The glaucoma patients were divided into early (MD≥-6dB, n=33) and moderate-to-advanced (MD<-6 dB, n=29) groups.
RESULTS
Significant correlations were found between the best-performing 4 RETeval ERG parameters and the glaucoma severity measures (MD and OCT thickness parameters) for all eyes, all glaucoma eyes and the moderate-to-advanced glaucoma eyes [photopic negative response amplitude at 72 ms (PhNR 72) and
MD
r=-0.333, -0.414, and -0.485, respectively, P≤0.008; PhNR 72 and average circumpapillary retinal nerve fiber layer thickness; r=-0.429, -0.450, and -0.542, respectively, P≤0.002]. Except for P-ratio, there was no significant difference between the area under the receiver-operating characteristic (AUROC) values of the OCT thickness parameters (range: 0.927 to 0.938) and the 4 best-performing RETeval ERG parameters (range: 0.839 to 0.905) in the early glaucoma versus control separation. For differentiating the control and the moderate-to-advanced glaucoma eyes, the AUROC values of the 4 best-performing RETeval ERG parameters ranged between 0.924 and 0.958, and no significant difference was found between them and those of the OCT parameters.
CONCLUSIONS
The noninvasive, portable RETeval full-field ERG device may be useful to detect glaucoma in moderate-to-advanced stages.
More by Yoshiyuki Kita
View full profile →Effect of Pilocarpine on the Iridocorneal Angle Following Phaco-Kahook Dual Blade Goniotomy in Japanese Glaucoma Patients.
Influence of Refractive Error on Circumpapillary Structure-Function Versus Vessel Density-Function Relationships in Open Angle Glaucoma.
Optical coherence tomographic findings of glaucomatous eyes with papillomacular retinoschisis.
Top Research in Visual Field
Browse all →Optical coherence tomography angiography: A comprehensive review of current methods and clinical applications.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Improving our understanding, and detection, of glaucomatous damage: An approach based upon optical coherence tomography (OCT).
In the Knowledge Library
Discussion
Comments and discussion will appear here in a future update.