Matching Misaligned Spectralis OCTs to a Reference Scan in Pediatric Glaucoma with Poor Fixation and Nystagmus.
Summary
Retinal layer thickness can be compared to a reference cpRNFL OCT scan when poor fixation and nystagmus causes misalignment errors. The analysis can be performed post hoc using multiple misaligned scans from standard SLO/OCT protocols.
Abstract
PURPOSE
Poor fixation or nystagmus in children causes misalignment errors when measuring circumpapillary retinal nerve fiber layer (cpRNFL) thickness by simultaneous scanning laser ophthalmoscope imaging/optical coherence tomography (SLO/OCT). We investigated a method to assess cpRNFL from misaligned SLO/OCT scans.
METHODS
Heidelberg Spectralis SLO/OCT scans from a single clinical examination were retrospectively analyzed when automated eye tracking was unreliable. Retinal layer thickness was measured at overlapping match locations between a reference and misaligned scans based on the position data from simultaneously acquired SLO images. Three layers were segmented: cpRNFL, internal limiting membrane to outer nuclear layer (ILM-ONL), and total retinal thickness (TR). Accuracy was defined as the difference in thickness between the reference and misaligned scans at their match locations after correction for scan angle.
RESULTS
Thirty-five subjects, evaluated for glaucomatous nerve loss, met inclusion criteria. Group-averaged accuracy was -2.7, 1.4, and 0.3 µm for cpRNFL, ILM-ONL, and TR thickness, respectively. Across all layers, interobserver intraclass correlation coefficients ranged from 0.97 to 0.63 and the maximum Bland-Altman 95% limits of agreement were -21.6 to 20.7 µm. Variability was greatest for cpRNFL thickness and least for TR thickness. Increased variability was associated with lower signal-to-noise ratio but not with image-motion indices of shear, rotation, and scale.
CONCLUSIONS
Retinal layer thickness can be compared to a reference cpRNFL OCT scan when poor fixation and nystagmus causes misalignment errors. The analysis can be performed post hoc using multiple misaligned scans from standard SLO/OCT protocols.
TRANSLATIONAL RELEVANCE
Our method allows for assessment of cpRNFL in children who fail eye tracking.
Keywords
Top Research in Optic Nerve & Disc
Browse all →Efficacy of a Deep Learning System for Detecting Glaucomatous Optic Neuropathy Based on Color Fundus Photographs.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Inflammation in Glaucoma: From the back to the front of the eye, and beyond.
In the Knowledge Library
Discussion
Comments and discussion will appear here in a future update.