False Positive Responses in Standard Automated Perimetry.
Anders Heijl, Vincent Michael Patella, John G Flanagan, Aiko Iwase, Christopher K Leung, Anja Tuulonen, Gary C Lee, Thomas Callan, Boel Bengtsson
Summary
Across 3 different standard automated perimetry thresholding algorithms, FP rates showed only weak associations with visual field test results, except in severe glaucoma.
Abstract
PURPOSE
To analyze the relationship between rates of false positive (FP) responses and standard automated perimetry results.
DESIGN
Prospective multicenter cross-sectional study.
METHODS
One hundred twenty-six patients with manifest or suspect glaucoma were tested with Swedish Interactive Thresholding Algorithm (SITA) Standard, SITA Fast, and SITA Faster at each of 2 visits. We calculated intervisit differences in mean deviation (MD), visual field index (VFI), and number of statistically significant test points as a function of FP rates and also as a function of general height (GH).
RESULTS
Increasing FP values were associated with higher MD values for all 3 algorithms, but the effects were small, 0.3 dB to 0.6 dB, for an increase of 10 percentage points of FP rate, and for VFI even smaller (0.6%-1.4%). Only small parts of intervisit differences were explained by FP (rvalues 0.00-0.11). The effects of FP were larger in severe glaucoma, with MD increases of 1.1 dB to 2.0 dB per 10 percentage points of FP, and r values ranging from 0.04 to 0.33. The numbers of significantly depressed total deviation points were affected only slightly, and pattern deviation probability maps were generally unaffected. GH was much more strongly related to perimetric outcomes than FP.
CONCLUSIONS
Across 3 different standard automated perimetry thresholding algorithms, FP rates showed only weak associations with visual field test results, except in severe glaucoma. Current recommendations regarding acceptable FP ranges may require revision. GH or other analyses may be better suited than FP rates for identifying unreliable results in patients who frequently press the response button without having perceived stimuli.
Keywords
More by Anders Heijl
View full profile →A New SITA Perimetric Threshold Testing Algorithm: Construction and a Multicenter Clinical Study.
Screening for Open-Angle Glaucoma and Its Effect on Blindness.
Long-term Impact of Immediate Versus Delayed Treatment of Early Glaucoma: Results From the Early Manifest Glaucoma Trial.
Top Research in Visual Field
Browse all →Optical coherence tomography angiography: A comprehensive review of current methods and clinical applications.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Improving our understanding, and detection, of glaucomatous damage: An approach based upon optical coherence tomography (OCT).
Discussion
Comments and discussion will appear here in a future update.