Virtual Reality-Based and Conventional Visual Field Examination Comparison in Healthy and Glaucoma Patients.
Summary
This study demonstrates that a clinically used perimeter and the proposed VR perimetry system have comparable performances with respect to a number of perimetry parameters in healthy and glaucoma patients with early to moderate visual…
Abstract
PURPOSE
Clinically evaluate the noninferiority of a custom virtual reality (VR) perimetry system when compared to a clinically and routinely used perimeter on both healthy subjects and glaucoma patients.
METHODS
We use a custom-designed VR perimetry system tailored for visual field testing. The system uses Oculus Quest VR headset (Facebook Technologies, LLC, Bern, Switzerland), that includes a clicker for participant response feedback. A prospective, single center, study was conducted at the Department of Ophthalmology of the Bern University Hospital (Bern, Switzerland) for 12 months. Of the 114 participants recruited 70 subjects (36 healthy and 34 glaucoma patients with early to moderate visual field loss) were included in the study. Participants underwent perimetry tests on an Octopus 900 (Haag-Streit, Köniz, Switzerland) as well as on the custom VR perimeter. In both cases, standard dynamic strategy (DS) was used in conjunction with the G testing pattern. Collected visual fields (VFs) from both devices were then analyzed and compared.
RESULTS
High mean defect (MD) correlations between the two systems (Spearman, ρ ≥ 0.75) were obtained. The VR system was found to slightly underestimate VF defects in glaucoma subjects (1.4 dB). No significant bias was found with respect to eccentricity or subject age. On average, a similar number of stimuli presentations per VF was necessary when measuring glaucoma patients and healthy subjects.
CONCLUSIONS
This study demonstrates that a clinically used perimeter and the proposed VR perimetry system have comparable performances with respect to a number of perimetry parameters in healthy and glaucoma patients with early to moderate visual field loss.
TRANSLATIONAL RELEVANCE
This suggests that VR perimeters have the potential to assess VFs with high enough confidence, whereby alleviating challenges in current perimetry practices by providing a portable and more accessible visual field test.
More by Jan Stapelfeldt
View full profile →Top Research in Visual Field
Browse all →Optical coherence tomography angiography: A comprehensive review of current methods and clinical applications.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Improving our understanding, and detection, of glaucomatous damage: An approach based upon optical coherence tomography (OCT).
In the Knowledge Library
Discussion
Comments and discussion will appear here in a future update.