Increased Depth, Reduced Extent, and Sharpened Edges of Visual Field Defects Measured by Compass Fundus Perimeter Compared to Humphrey Field Analyzer.

Visual FieldOptic Nerve & Disc

Summary

Glaucomatous visual fields measured by CMP displayed characteristics consistent with expected effects of using real-time eye movement compensation technology compared to the widely used HFA.

Abstract

PURPOSE

The purpose of this study was to compare visual field results of the COMPASS fundus perimeter (CMP) and the Humphrey Field Analyzer (HFA) in the same eyes; to compare structure-function concordance between circumpapillary retinal nerve fiber layer (Cp-RNFL) profiles and the two perimetry results; and to evaluate whether differences between the two results reflect postulated advantages of real-time eye movement compensation during perimetry.

METHODS

We retrospectively analyzed 24-2 visual field data measured with CMP and HFA together with Cp-RNFL optical coherence tomography (OCT) scan data from 95 eyes of 65 people with glaucoma. We defined visual field locations with total deviation (TD) less than -5 dB as defective. The CMP and HFA fields were compared on measures of: spatial extent (number of defective locations); depth (TD values); and sharpness of scotomata edges (maximum TD difference between defective locations and their neighbors). Structure-function concordance between Cp-RNFL profile and respective visual field was also compared.

RESULTS

Compared to the HFA, scotomata measured by CMP were of reduced spatial extent (mean difference = -3.14 locations, p < 0.001), greater depth (median TD of CMP = -17 dB versus HFA = -13 dB, p = 0.029) and steeper edges (median of maximum TD difference of CMP = 10.6 dB versus HFA = 6 dB, p < 0.001). Structure-function concordance between Cp-RNFL profile and either visual field were comparable despite the reduced scotoma spatial extent measured by CMP.

CONCLUSIONS

Glaucomatous visual fields measured by CMP displayed characteristics consistent with expected effects of using real-time eye movement compensation technology compared to the widely used HFA.

TRANSLATIONAL RELEVANCE

Glaucomatous visual field defects measured by the CMP are more localized, deeper, and steeper than those of the HFA.