Three-Year Outcomes of a Schlemm Canal Microstent (Hydrus Microstent) with Concomitant Phacoemulsification in Open-Angle Glaucoma.
Summary
The 36-month results from this study show that the Hydrus microstent with phacoemulsification is safe and effective in reducing the IOP and AGM among patients with mild to severe open-angle glaucoma and can slow down…
Abstract
PURPOSE
HORIZON trials have reported the medium- and long-term outcomes of the Hydrus microstent. However, outside of randomized clinical trials, outcomes beyond 24 months have not been previously reported. Here, we examine the 3-year outcomes of the Hydrus microstent with cataract surgery outside of a randomized clinical trial setting.
DESIGN
Single-surgeon consecutive case series.
PARTICIPANTS
Glaucomatous eyes that underwent implantation of a Hydrus microstent with concomitant cataract surgery at a single ophthalmology center in Montreal, Canada.
METHODS
Success was defined according to the absence of specific failure criteria: (A) glaucoma reoperation; (B) selective laser trabeculoplasty; (C) intraocular pressure (IOP) 18 mmHg, or increase in the number of antiglaucoma medications (AGMs) used (after the first postoperative month), or loss of light perception due to glaucoma; (D) aggregation of criteria A-C. Predictors of treatment failure and postoperative changes in IOP and AGM use were assessed. Safety included best-corrected visual acuity (BCVA), cup-to-disc ratio (CDR), retinal nerve fiber layer (RNFL) thickness, ganglion cell inner plexiform layer (GC-IPL) thickness, visual field mean deviation (VF-MD) and pattern standard deviation (VF-PSD), and adverse events.
MAIN OUTCOME MEASURES
Surgical success, predictors of treatment failure, postoperative changes in IOP and AGM use, and structural and functional markers of disease stability.
RESULTS
One hundred six eyes were included. The 3-year surgical success according to Criteria A to D was 86%, 83%, 91%, and 67%, respectively. Intraocular pressure decreased by 26.5% from 18.9 ± 4.8 to 13.9 ± 2.3 mmHg (P 0.05). Postoperative adverse events were few and transient.
CONCLUSIONS
The 36-month results from this study show that the Hydrus microstent with phacoemulsification is safe and effective in reducing the IOP and AGM among patients with mild to severe open-angle glaucoma and can slow down the disease progression by preserving both structural and functional parameters. FINANCIAL DISCLOSURE(S): Proprietary or commercial disclosure may be found after the references.
Keywords
More by Ali Salimi
View full profile →Gonioscopy-Assisted Transluminal Trabeculotomy in Younger to Middle-Aged Adults: One-Year Outcomes.
Three-Year Outcomes of Second-generation Trabecular Micro-bypass Stents (iStent inject) With Phacoemulsification in Various Glaucoma Subtypes and Severities.
Matched Cohort Study of Cataract Surgery With and Without Trabecular Microbypass Stent Implantation in Primary Angle-Closure Glaucoma.
Top Research in Optic Nerve & Disc
Browse all →Efficacy of a Deep Learning System for Detecting Glaucomatous Optic Neuropathy Based on Color Fundus Photographs.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Inflammation in Glaucoma: From the back to the front of the eye, and beyond.
Discussion
Comments and discussion will appear here in a future update.