Comparative Analysis of 10-2 Test on Advanced Vision Analyzer and Humphrey Perimeter in Glaucoma.
Summary
The Advanced Vision Analyzer effectively detects glaucoma, showing strong correlation and comparable diagnostic precision to the Humphrey Field Analyzer for 10-2 visual field testing.
Abstract
PURPOSE
To evaluate diagnostic precision and prove equivalence of 2 devices, Advanced vision analyzer (AVA, Elisar Vision Technology) and Humphrey field analyzer (HFA, Zeiss) for the detection of glaucoma on 10-2 program.
DESIGN
Prospective, cross-sectional, observational study.
PARTICIPANTS
Threshold estimates of 1 eye each of 66 patients with glaucoma, 36 control participants, and 10 glaucoma suspects were analyzed on 10-2 test with AVA and HFA.
METHODS
Mean sensitivity (MS) values of 68 points and central 16 test points were calculated and compared. Intraclass correlation (ICC), Bland-Altman (BA) plots, linear regression of MS, mean deviation (MD), and pattern standard deviation (PSD) were computed to assess the 10-2 threshold estimate of the devices. Receiver operating characteristic curves were generated for MS and MD values, and the area under the curve (AUC) was compared with assessing diagnostic precision.
MAIN OUTCOME MEASURES
Mean sensitivity values of 68 points and central 16 points, AUC for MS and MD values, ICC values, BA plots, and linear-regression analysis.
RESULTS
Bland-Altman plot showed significant correlation for MS, MD, and PSD values for both devices. For MS, the overall ICC value was 0.96 ( 0.05). The AUC for MS values for AVA was 0.89 and for HFA was 0.92 ( = 0.188); whereas it was similar at 0.88 for MD values ( = 0.799). Advanced vision analyzer and HFA identically discriminated between healthy and patients with glaucoma ( 0.05).
CONCLUSIONS
Statistical results denote adequate equivalence between AVA and HFA because threshold estimates of AVA strongly correlate with HFA for 10-2 program.
FINANCIAL DISCLOSURES
Proprietary or commercial disclosure may be found after the references.
Keywords
More by Priya Narang
View full profile →Comparison of Elisar-Fast and Sita-Fast Strategies for Visual Field Assessment in Glaucoma.
November consultation #3.
Advanced Vision Analyzer-Virtual Reality Perimeter: Device Validation, Functional Correlation and Comparison with Humphrey Field Analyzer.
Top Research in Visual Field
Browse all →Optical coherence tomography angiography: A comprehensive review of current methods and clinical applications.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Improving our understanding, and detection, of glaucomatous damage: An approach based upon optical coherence tomography (OCT).
This article has not yet been placed in the Knowledge Library.
Discussion
Comments and discussion will appear here in a future update.