Validation of the Iowa Head-Mounted Open-Source Perimeter.
Zachary Heinzman, Edward Linton, Iván Marín-Franch, Andrew Turpin, Karam Alawa, Anushi Wijayagunaratne, Michael Wall
Summary
The Iowa Virtual Reality HMD Open-Source Perimeter is as repeatable as the Octopus 900 perimeter and is a more portable and less expensive alternative than traditional perimeters.
Abstract
PURPOSE
To assess the validity of visual field (VF) results from the Iowa Head-Mounted Display (HMD) Open-Source Perimeter and to test the hypothesis that VF defects and test-retest repeatability are similar between the HMD and Octopus 900 perimeters.
METHODS
We tested 20 healthy and nine glaucoma patients on the HMD and Octopus 900 perimeters using the Open Perimetry Interface platform with size V stimuli, a custom grid spanning the central 26° of the VF, and a ZEST thresholding algorithm. Historical data from the Humphrey Field Analyzer (HFA) were also analyzed. Repeatability was analyzed with the repeatability coefficient (RC), and VF defect detection was determined through side-by-side comparisons.
RESULTS
The pointwise RCs were 2.6 dB and 3.4 dB for the HMD and Octopus 900 perimeters in ocular healthy subjects, respectively. Likewise, the RCs were 4.2 dB and 3.5 dB, respectively, in glaucomatous patients. Limits of agreement between the HMD and Octopus 900 perimeters were ±4.6 dB (mean difference, 0.4 dB) for healthy patients and ±8.9 dB (mean difference, 0.1 dB) for glaucomatous patients. Retrospective analysis showed that pointwise RCs on the HFA2 perimeter were between 3.4 and 3.7 dB for healthy patients and between 3.9 and 4.7 dB for glaucoma patients. VF defects were similar between the HMD and Octopus 900 for glaucoma subjects.
CONCLUSIONS
The Iowa Virtual Reality HMD Open-Source Perimeter is as repeatable as the Octopus 900 perimeter and is a more portable and less expensive alternative than traditional perimeters.
TRANSLATIONAL RELEVANCE
This study demonstrates the validity of the visual field results from the Iowa HMD Open-Source Perimeter which may help expand perimetry access.
Top Research in Visual Field
Browse all →Optical coherence tomography angiography: A comprehensive review of current methods and clinical applications.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Improving our understanding, and detection, of glaucomatous damage: An approach based upon optical coherence tomography (OCT).
In the Knowledge Library
Discussion
Comments and discussion will appear here in a future update.