Measuring Visual Fields in Children With Glaucoma Using a Portable Tablet.
Viney Gupta, George X Y Kong, Abhishek Singh, Arnav Panigrahi, Shikha Gupta, Selwyn Prea, Algis J Vingrys
Summary
MRF correlated strongly with HFA and was preferred by the children over the HFA. MRF is suitable for perimetric evaluation of children with glaucoma.
Abstract
PURPOSE
To compare perimetric outcomes of an iPad perimetry app (Melbourne Rapid Fields [MRF]) with those of the Humphrey Field Analyser (HFA) testing children with glaucoma.
METHODS
Sixteen children diagnosed and treated for glaucoma were recruited to evaluate their perimetric performance over two visits. At each visit, they undertook visual field assessment using the MRF application as well as the HFA. The HFA test was part of their usual clinical work up and a clinical assistant judged which test format (24-2 SITA standard or SITA fast) might be suited to the testing of that child. The primary outcome measure was the association and repeatability of mean deviation (MD) for the MRF and HFA tests, by way of regression, intraclass correlation coefficient and Bland-Altman analysis. Secondary measures were comparisons of pattern deviation indices, test times as well as an indication of participant test preference. Summary data show means ± standard deviation.
RESULTS
The age for our cohort was 7 to 15 years of age (mean, 10.0 ± 2.4 years of age). The MRF MD was in close concordance to HFA MD with an intraclass correlation coefficient of 0.91 (95% confidence interval, 0.82-0.95). Bland-Altman analysis found little bias (-0.6 dB) and a 95% coefficient of repeatability of 2.1 dB in eyes having a normal HFA MD. In eyes with glaucomatous visual field defects the 95% coefficient of repeatability at retest was much larger for both the MRF (10.5 dB) as well as for the HFA (10.0 dB). Average MRF test times (5.6 ± 1.2 minutes) were similar to SITA Fast (5.4 ± 1.9 minutes) with both being significantly faster than SITA standard (8.6 ± 1.4 minutes; P < 0.001). All children chose testing with the MRF as their preference.
CONCLUSIONS
MRF correlated strongly with HFA and was preferred by the children over the HFA. MRF is suitable for perimetric evaluation of children with glaucoma.
TRANSLATIONAL RELEVANCE
This study finds that an iPad based visual field test can be used with children having glaucoma to yield outcomes similar to SITA-fast. Children indicate a preference for such testing.
More by Viney Gupta
View full profile →Long-term outcomes of glaucoma drainage devices for glaucoma post-vitreoretinal surgery with silicone oil insertion: a prospective evaluation.
Safety and efficacy of incisional goniotomy as an adjunct with phacoemulsification in primary angle-closure glaucoma.
In Vivo Analysis of Angle Dysgenesis in Primary Congenital, Juvenile, and Adult-Onset Open Angle Glaucoma.
Top Research in Visual Field
Browse all →Optical coherence tomography angiography: A comprehensive review of current methods and clinical applications.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Improving our understanding, and detection, of glaucomatous damage: An approach based upon optical coherence tomography (OCT).
In the Knowledge Library
Discussion
Comments and discussion will appear here in a future update.