Glaucoma Home Self-Testing Using VR Visual Fields and Rebound Tonometry Versus In-Clinic Perimetry and Goldmann Applanation Tonometry: A Pilot Study.
Summary
Unsupervised, at-home, multi-day glaucoma testing using two devices resulted in the capture of higher maximum IOPs than in the clinic and good MD correlation of VRP with HFA.
Abstract
PURPOSE
This pilot study aimed to assess the feasibility, accuracy, and repeatability of unsupervised, at-home, multi-day glaucoma testing using the Olleyes VisuALL Virtual Reality Platform (VRP) and the iCare HOME handheld self-tonometer.
METHODS
Participants were trained to use two U.S. Food and Drug Administration-registered or approved devices before conducting self-tests at home over 3 consecutive days. The iCare HOME intraocular pressure (IOP) measurements were collected four times daily per eye, and VRP visual field tests were performed once daily. The results were compared with one in-clinic Humphrey Field Analyzer (HFA) visual field test performed on the day of device training, iCare HOME measurements by the trainer, and the last five Goldmann applanation tonometer (GAT) results.
RESULTS
Of 15 enrolled participants, nine of them (60%) completed the study. The six excluded participants could not self-measure using iCare HOME. There was significant correlation between the average mean deviation (MD) values of the at-home VRP tests and in-clinic HFA test (r2 = 0.8793, P < 0.001). Additionally, the average of the sensitivities in five of six Garway-Heath sectors were significantly correlated. VRP test duration was also shorter than in-clinic HFA testing (P < 0.001). Finally, at-home tonometry yielded statistically similar values compared to trainer-obtained iCare HOME values. The mean and range of at-home tonometry were also statistically similar to those for in-clinic GAT, but at-home tonometry demonstrated higher maximum IOP values (P = 0.0429).
CONCLUSIONS
Unsupervised, at-home, multi-day glaucoma testing using two devices resulted in the capture of higher maximum IOPs than in the clinic and good MD correlation of VRP with HFA. However, 40% of participants could not self-measure IOP using iCare HOME.
TRANSLATIONAL RELEVANCE
The study findings suggest that at-home remote glaucoma monitoring correlates with in-office testing and could provide additional information for glaucoma management, although patients had more difficulty with the iCare HOME than the VRP.
More by Andrew R Berneshawi
View full profile →Top Research in Visual Field
Browse all →Optical coherence tomography angiography: A comprehensive review of current methods and clinical applications.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Improving our understanding, and detection, of glaucomatous damage: An approach based upon optical coherence tomography (OCT).
In the Knowledge Library
Discussion
Comments and discussion will appear here in a future update.