Comparison of the Humphrey Field Analyzer and MP-3 Microperimeter in Patients With Glaucoma, Classified by Severity and Misalignment of Test Points.
Summary
Reliability indices differed significantly between the two devices in severe POAG cases. A shift in the sensitivity of the test points of the two VF tests was detected in approximately one-half of the eyes.
Abstract
PURPOSE
To compare Nidek MP-3 microperimetry and ZEISS Humphrey Field Analyzer (HFA) visual field (VF) results in patients with primary open-angle glaucoma (POAG), classified by VF defect severity and to describe a novel method for simulating sensitivity distribution changes assumed to be due to fixation errors.
METHODS
This retrospective study used the MP-3 and HFA 10-2 tests to examine the VF in eyes with mild, moderate, or severe POAG, using 68 identical test points. Reliability indices and measurement durations were compared. The relationship between pointwise sensitivities of the devices was analyzed.
RESULTS
Thirty-five eyes from 23 patients (10 mild, 12 moderate, and 13 severe POAG) were evaluated. In the severe POAG group, the MP-3 had a significantly lower false-positive (FP) rate and higher false-negative rate than those of the HFA 10-2 (all, P = 0.03). A significant negative correlation existed between the root mean square error, derived from regression analysis comparing the VF sensitivity between devices at each test point, and VF sensitivity of the MP-3 (P < 0.01). In 18 eyes, a shift in the sensitivity distribution occurred, with a significant correlation between the HFA 10-2 mean deviation and the presence of this shift (P < 0.01).
CONCLUSIONS
Reliability indices differed significantly between the two devices in severe POAG cases. A shift in the sensitivity of the test points of the two VF tests was detected in approximately one-half of the eyes.
TRANSLATIONAL RELEVANCE
The MP-3 provides lower FP rates in advanced POAG and may assist in the management of this patient cohort.
More by Taiga Inooka
View full profile →Top Research in Visual Field
Browse all →Optical coherence tomography angiography: A comprehensive review of current methods and clinical applications.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Improving our understanding, and detection, of glaucomatous damage: An approach based upon optical coherence tomography (OCT).
In the Knowledge Library
Discussion
Comments and discussion will appear here in a future update.