Residual Iris Retraction Syndrome After Artificial Iris Implantation.
Mayer Christian S, Laubichler Andrea E, Masyk Michael, Prahs Philipp, Zapp Daniel, Khoramnia Ramin
AI Summary
Artificial iris implantation can cause residual iris retraction syndrome, where the remaining iris dilates and doesn't constrict properly, potentially leading to complications.
Abstract
Purpose
To evaluate the effect of an artificial iris implant on the remnant iris.
Design
Interventional case series.
Methods
Setting: Single center.
Patient population: Forty-two consecutive patients.
Observation procedures: Morphologic evaluation over 24 ± 14 months.
Main outcome measures
Remnant pupillary aperture, iris color, visual acuity, intraocular pressure, and endothelial cell count.
Results
In 7 of 42 cases (16.7%), the residual iris aperture dilated from 36.6 ± 15.4 mm 2 preoperatively to 61.1 ± 12.5 mm 2 1 year postoperatively (66.9% increase). In 5 of 7 affected eyes the artificial iris had been implanted into the ciliary sulcus; in 2 eyes it had been sutured to the sclera. Four of the 7 patients presented with remarkable complications: 2 eyes needed glaucoma shunt surgeries owing to pigment dispersion; 1 suffered from recurrent bleedings; and in 1 case artificial iris explantation was performed owing to chronic inflammation. Anterior chamber depth and angle, endothelial cell count, and visual acuity did not change in this cohort. Changes in color were not observed in the remnant iris.
Conclusions
The implantation of an artificial iris prosthesis can lead to a residual iris retraction syndrome. It is likely that residual iris is trapped in the fissure between the artificial iris and the anterior chamber angle, preventing further pupil constriction. Another possibility could be a constriction or atrophy of the residual iris. A scleral-sutured implant and an implantation in the capsular bag were both found to prevent the iris retraction. The study group number is inadequate to allow statistical comparison of these different implantation methods. As the use of artificial irises increases, we may expect more patients with iris retraction syndrome in the future.
MeSH Terms
Shields Classification
Key Concepts5
The implantation of an artificial iris prosthesis can lead to a residual iris retraction syndrome.
In 7 of 42 cases (16.7%) of artificial iris implantation, the residual iris aperture dilated from 36.6 15.4 mm2 preoperatively to 61.1 12.5 mm2 1 year postoperatively (66.9% increase), indicating residual iris retraction syndrome.
Among 7 patients who developed residual iris retraction syndrome after artificial iris implantation, 4 presented with remarkable complications: 2 eyes needed glaucoma shunt surgeries due to pigment dispersion, 1 suffered from recurrent bleedings, and in 1 case artificial iris explantation was performed due to chronic inflammation.
A scleral-sutured artificial iris implant and an artificial iris implantation in the capsular bag were both found to prevent residual iris retraction in a case series of 42 patients.
Anterior chamber depth and angle, endothelial cell count, and visual acuity did not change in a cohort of 7 patients who developed residual iris retraction syndrome after artificial iris implantation.
Related Articles5
Complications and Management of Cosmetic Iris Implants.
Case SeriesAqueous misdirection syndrome: clinical outcomes and risk factors for treatment failure.
Retrospective StudyIritis, glaucoma and corneal decompensation associated with BrightOcular cosmetic iris implant.
Case SeriesPersistent leak after glaucoma aqueous shunt implantation.
Case SeriesA traumatic "peripheral iridotomy" protects against pigment dispersion and glaucoma.
Case ReportIs this article assigned to the wrong chapter(s)? Let us know.