A Pilot Study on Feasibility and Effectiveness of Intraoperative Spectral-Domain Optical Coherence Tomography in Glaucoma Procedures.
Rajesh S Kumar, Manan U Jariwala, Sathidevi A V, Jayasree P Venugopal, Narendra K Puttaiah, Ramgopal Balu, A S Dhanaraj Rao, Rohit Shetty
Summary
TheOCT system provided high quality, intraoperative, real-time imaging, which could possibly improve the safety and efficacy of the surgical procedures in glaucoma.
Abstract
PURPOSE
To demonstrate the use of a spectral-domain optical coherence tomography (SDOCT) integrated surgical microscope in glaucoma surgery.
METHODS
An SDOCT system was used to interface directly with an ophthalmic surgical microscope, to allow real-time intraoperative SDOCT (OCT) imaging during glaucoma procedures like phaco-trabeculectomy, Ahmed glaucoma valve (AGV) implantation, gonio-synechiolysis, and bleb needling. The various surgical steps during glaucoma surgeries whereOCT can be of potential help in guiding the surgeon were recorded.
RESULTS
High-resolution, cross-sectional images of the relevant structures were achieved with theOCT system in all procedures. The surgeon could determine the depth of the scleral dissection, the intrastomal bed, the path of the AGV tube in the eye, the release of peripheral anterior synechiae and the efficacy of needling with respect to breakage of loculations; most of these are technically 'blind' procedures, where the outcomes are determined postoperatively. Metallic instruments cast a shadow on tissues below, thereby restricting the use of the device in its current state.
CONCLUSIONS
TheOCT system provided high quality, intraoperative, real-time imaging, which could possibly improve the safety and efficacy of the surgical procedures in glaucoma. Further studies and modifications to theOCT are required to better understand and increase the uptake of this technology in daily practice.
TRANSLATIONAL RELEVANCE
TheOCT, with further advancements in its technology, could potentially provide the surgeon both quantitative and qualitative, real-time depth and tissue proximity details, thus improving the safety and accuracy of glaucoma surgery.
Keywords
More by Rajesh S Kumar
View full profile →Diagnostic ability of peripapillary vessel density measurements of optical coherence tomography angiography in primary open-angle and angle-closure glaucoma.
Discriminant Function of Optical Coherence Tomography Angiography to Determine Disease Severity in Glaucoma.
The Singapore Asymptomatic Narrow Angles Laser Iridotomy Study: Five-Year Results of a Randomized Controlled Trial.
Top Research in Glaucoma Surgery
Browse all →Treatment Outcomes in the Primary Tube Versus Trabeculectomy Study after 1 Year of Follow-up.
Efficacy, Safety, and Risk Factors for Failure of Standalone Ab Interno Gelatin Microstent Implantation versus Standalone Trabeculectomy.
Prospective, Randomized, Controlled Pivotal Trial of an Ab Interno Implanted Trabecular Micro-Bypass in Primary Open-Angle Glaucoma and Cataract: Two-Year Results.
Discussion
Comments and discussion will appear here in a future update.