Corneal Deformation Response in Patients With Primary Open-Angle Glaucoma and in Healthy Subjects Analyzed by Corvis ST.
Summary
The corneal response parameters provided by CST are informative for the assessment of corneal biomechanics.
Abstract
PURPOSE
To compare corneal deformation response between patients with primary open-angle glaucoma (POAG) and healthy subjects. A prevalent case-control study was conducted, followed by an integrated meta-analysis.
METHODS
Primary open-angle glaucoma was confirmed by presence of glaucomatous optic disc damage with corresponding visual field defect after excluding secondary causes. Age-matched controls were recruited. Corvis ST (CST) was used to measure 10 parameters of corneal deformation response and central corneal thickness (CCT). The association between parameters of deformation response and clinical factors was assessed by the linear regression analyses. Differences in deformation response between POAG and healthy subjects were compared after adjusting for other factors. Eligible studies were identified by a systematic search of the PubMed, ISI Web of Science, and Embase databases; Web sites of professional associations; and the Google Scholar engine.
RESULTS
This study included 37 patients with POAG and 36 healthy controls. Significant associations were found between IOP and the first applanation time (A1T), first applanation velocity (A1V), second applanation time (A2T), and second applanation velocity (A2V) in each group. The deformation amplitude (DA), A1V, and A2T were lower in the POAG group, whereas the A1T, A2V and peak distance (PD) were greater in the POAG group than in the healthy controls (all P < 0.05). Ten separate studies plus the present study, involving 691 patients with POAG and 740 controls, were ultimately meta-analyzed. The significant differences in the DA, A1T, and A2T were further confirmed, with pooled weighted mean difference (WMD) of 0.13 mm (95% CI 0.04-0.23; P = 0.008) for A1T, -0.13 mm (95% CI -0.21 to -0.05; P = 0.001) for DA, and -0.46 ms (95% CI -0.70 to -0.22; P < 0.001) for A2T.
CONCLUSIONS
The corneal response parameters provided by CST are informative for the assessment of corneal biomechanics. Patients with POAG showed significantly greater A1T and lower DA and A2T values than healthy controls, indicating a less deformable cornea in POAG patients.
More by Wei Wang
View full profile →Epidemiological variations and trends in health burden of glaucoma worldwide.
Classification of Visual Field Abnormalities in Highly Myopic Eyes without Pathologic Change.
Association of Long-term Ambient Black Carbon Exposure and Oxidative Stress Allelic Variants With Intraocular Pressure in Older Men.
Top Research in Cornea & Biomechanics
Browse all →The impact of chronic use of prostaglandin analogues on the biomechanical properties of the cornea in patients with primary open-angle glaucoma.
Corneal Biomechanics and Visual Field Progression in Eyes with Seemingly Well-Controlled Intraocular Pressure.
Evaluation of a New Rebound Self-tonometer, Icare HOME: Comparison With Goldmann Applanation Tonometer.
Discussion
Comments and discussion will appear here in a future update.