Interobserver agreement using Goldmann applanation tonometry and dynamic contour tonometry: comparing ophthalmologists, nurses and technicians.
Summary
Ophthalmologists show good levels of agreement with each other when using GAT, while technicians/nursing staff show better agreement when using the DCT.
Abstract
PURPOSE
To compare agreement between ophthalmologists and non-ophthalmologists (nurses and ophthalmic technicians) when measuring the intraocular pressure (IOP) using Goldmann applanation tonometry (GAT) and Pascal dynamic contour tonometry (DCT).
METHODS
Patients attending for their routine glaucoma outpatient appointment were invited to participate. IOP was measured in one eye by either two ophthalmologists (observer group 1), two non-ophthalmologists (observer group 2) or one ophthalmologist and one non-ophthalmologist (observer group 3). All were experienced in using GAT and some experienced in using the DCT. The order of tonometer and staff was randomised. Agreement was calculated by Bland-Altman analysis, with the mean difference and 95% limits of agreement (LoA) of measurements calculated for each observer group.
RESULTS
One hundred eyes were measured within each observer group. The mean difference (95% LoA) in IOP measurements were
GAT
group 1=-0.20 (4.9) mm Hg, group 2=0.6 (5.4) mm Hg and group 3=0.0 (3.7) mm Hg;
DCT
group 1=0.8 (7.7) mm Hg, group 2=0.3 (4.2) and group 3=0.0 (5.2) mm Hg. The DCT consistently over-read the GAT for all observer groups.
DISCUSSION/CONCLUSIONS
Ophthalmologists show good levels of agreement with each other when using GAT, while technicians/nursing staff show better agreement when using the DCT. The DCT may be a better tonometer to use if permanently delegating IOP measurements to non-ophthalmologists, but measurements cannot be interchanged with the GAT.
Keywords
More by Aachal Kotecha
View full profile →Virtual clinics in glaucoma care: face-to-face versus remote decision-making.
Quality of Life in the Tube Versus Trabeculectomy Study.
Developing standards for the development of glaucoma virtual clinics using a modified Delphi approach.
Top Research in IOP & Medical Therapy
Browse all →The Complications of Myopia: A Review and Meta-Analysis.
Inflammation in Glaucoma: From the back to the front of the eye, and beyond.
Treatment Outcomes in the Primary Tube Versus Trabeculectomy Study after 1 Year of Follow-up.
Discussion
Comments and discussion will appear here in a future update.