Invest Ophthalmol Vis Sci
Invest Ophthalmol Vis SciAugust 2016Multicenter Study

A Population-Based Assessment of the Agreement Between Grading of Goniophotographic Images and Gonioscopy in the Chinese-American Eye Study (CHES).

Angle & Aqueous OutflowIOP & Medical Therapy

Summary

Detection of iridocorneal angle closure based on goniophotographic imaging shows moderate to very good agreement with angle closure assessment using gonioscopy.

Abstract

PURPOSE

To compare grading of goniophotographic images and gonioscopy in assessing the iridocorneal angle.

METHODS

In a population-based, cross-sectional study, participants underwent gonioscopy and goniophotographic imaging during the same visit. The iridocorneal angle was classified as closed if the posterior trabecular meshwork could not be seen. A single masked observer graded the goniophotographic images, and each eye was classified as having angle closure based on the number of closed quadrants. Agreement between the methods was analyzed by calculating kappa (κ) and first-order agreement coefficient (AC1) statistics and comparison of area under receiver operating characteristic curves (AUC).

RESULTS

A total of 4149 Chinese Americans (3994 eyes) were included in this study. The agreement for angle closure diagnosis between gonioscopy and EyeCam was moderate to excellent (κ = 0.60, AC1 0.90, AUC 0.76-0.80).

CONCLUSIONS

Detection of iridocorneal angle closure based on goniophotographic imaging shows moderate to very good agreement with angle closure assessment using gonioscopy.