Interaction of primary human trabecular meshwork cells with metal alloy candidates for microinvasive glaucoma surgery.
Summary
Substratum topography and metal alloy composition were found to impact morphology, viability and function of primary HTMC cultures.
Abstract
BACKGROUND
Microinvasive glaucoma surgery (MIGS) is a relatively new addition to the glaucoma treatment paradigm. Small metallic stents are inserted into the trabecular meshwork in order to increase aqueous humour drainage. MIGS procedures are rapidly being adopted owing to a more favourable side effect profile when compared with traditional surgery. Remarkably, this rapid rate of utilization has occurred without any published studies on the effect of metal alloys used in these stents on human trabecular meshwork cells (HTMCs). Therefore, this study aimed to determine the effect of candidate metal alloys for MIGS on HTMC morphology, viability and function.
METHODS
Human trabecular meshwork cells were cultured on the surfaces of titanium (polished and sandblasted), a titanium-nickel (nitinol) alloy and glass (as control substratum). Fluorescence imaging was used to assess cell morphology and spreading. A lactate dehydrogenase cytotoxicity assay, cell death detection ELISA, MTT cell viability assay, BrdU cell proliferation assay and fibronectin ELISA were also conducted.
RESULTS
Cells cultured on sandblasted titanium exhibited significantly greater spreading than cells cultured on other substrata. In comparison, HTMCs cultured on nitinol displayed poor spreading. Significantly more cell death, by both necrosis and apoptosis, occurred on nitinol than on titanium and glass. Also, cell viability and proliferation were suppressed on nitinol compared with titanium or glass. Finally, HTMCs on both titanium and nitinol produced greater amounts of fibronectin than cells grown on glass.
CONCLUSIONS
Substratum topography and metal alloy composition were found to impact morphology, viability and function of primary HTMC cultures.
Keywords
Top Research in Glaucoma Surgery
Browse all →Treatment Outcomes in the Primary Tube Versus Trabeculectomy Study after 1 Year of Follow-up.
Efficacy, Safety, and Risk Factors for Failure of Standalone Ab Interno Gelatin Microstent Implantation versus Standalone Trabeculectomy.
Prospective, Randomized, Controlled Pivotal Trial of an Ab Interno Implanted Trabecular Micro-Bypass in Primary Open-Angle Glaucoma and Cataract: Two-Year Results.
Discussion
Comments and discussion will appear here in a future update.