Performance and Safety of a New Ab Interno Gelatin Stent in Refractory Glaucoma at 12 Months.
Davinder S Grover, William J Flynn, Kent P Bashford, Richard A Lewis, Yi-Jing Duh, Rupali S Nangia, Barbara Niksch
Summary
The gelatin stent reduced IOP and medication use without raising unexpected safety concerns, offering a minimally invasive surgical option for refractory glaucoma patients.
Abstract
PURPOSE
To evaluate the intraocular pressure (IOP)-lowering performance and safety of an ab interno gelatin stent (XEN 45 Gel Stent, Allergan plc, Irvine, California, USA), a minimally invasive glaucoma surgery device, in refractory glaucoma.
DESIGN
Single-arm, open-label, multicenter clinical study.
METHODS
Following mitomycin C pretreatment, the stent was placed ab interno in patients who failed prior filtering/cilioablative procedure or had uncontrolled IOP on maximum-tolerated medical therapy, with medicated IOP ≥20 and ≤35 mm Hg and visual field mean deviation ≤-3 dB. Primary performance outcomes: patients (%) achieving ≥20% IOP reduction from baseline on the same or fewer medications and mean IOP change from baseline at month 12. Procedure-related complications and ocular adverse events (AEs) were assessed.
RESULTS
Sixty-five patients were implanted (intent-to-treat/safety population). At 12 months, 75.4% (46/61; observed data) reported ≥20% IOP lowering from baseline on the same or fewer medications. Mean IOP change from baseline was -9.1 mm Hg (95% confidence interval [CI]: -10.7, -7.5) (n = 52; observed data) at 12 months, excluding patients with missing data (n = 4) and those requiring a glaucoma-related secondary surgical intervention (n = 9). Mean medication count decreased from 3.5 (baseline) to 1.7 (12 months). No intraoperative complications or unexpected postoperative AEs were reported. Most AEs were mild/moderate; common AEs included needling (without sight-threatening complications), nonpersistent loss of best-corrected visual acuity, and transient hypotony (requiring no surgical intervention).
CONCLUSIONS
The gelatin stent reduced IOP and medication use without raising unexpected safety concerns, offering a minimally invasive surgical option for refractory glaucoma patients.
More by Davinder S Grover
View full profile →Gonioscopy-assisted Transluminal Trabeculotomy: An Ab Interno Circumferential Trabeculotomy: 24 Months Follow-up.
Micropulse Cyclophotocoagulation: Initial Results in Refractory Glaucoma.
Ab-Externo MicroShunt versus Trabeculectomy in Primary Open-Angle Glaucoma: One-Year Results from a 2-Year Randomized, Multicenter Study.
Top Research in Glaucoma Surgery
Browse all →Treatment Outcomes in the Primary Tube Versus Trabeculectomy Study after 1 Year of Follow-up.
Efficacy, Safety, and Risk Factors for Failure of Standalone Ab Interno Gelatin Microstent Implantation versus Standalone Trabeculectomy.
Prospective, Randomized, Controlled Pivotal Trial of an Ab Interno Implanted Trabecular Micro-Bypass in Primary Open-Angle Glaucoma and Cataract: Two-Year Results.
Discussion
Comments and discussion will appear here in a future update.