Reproducibility of Optical Coherence Tomography Angiography Macular and Optic Nerve Head Vascular Density in Glaucoma and Healthy Eyes.
Patricia I C Manalastas, Linda M Zangwill, Luke J Saunders, Kaweh Mansouri, Akram Belghith, Min Hee Suh, Adeleh Yarmohammadi, Rafaella C Penteado, Tadamichi Akagi, Takuhei Shoji, Robert N Weinreb
Summary
Reproducibility of OCT-A ONH and macula vessel density measurements is good. Moreover, glaucoma patients have sparser vessel density with poorer reproducibility than healthy subjects.
Abstract
PURPOSE
Optical coherence tomography angiography (OCT-A) is a noninvasive technology that allows visualization of retinal blood vessels. It is important to determine reproducibility of measurements as low precision can impair its diagnostic capabilities. The purpose of this study is to determine intravisit and intervisit reproducibility of optic nerve head (ONH) and macular vessel density measurements with OCT-A.
PATIENTS AND METHODS
Fifteen healthy volunteers and 14 glaucoma patients completed 2 OCT-A (AngioVue; Optovue Inc.) scanning sessions on each of 2 separate days to assess intravisit and intervisit reproducibility. A series of ONH and macula scans were acquired at each session. Vessel density (%), the proportion of vessel area over the total measurement area was calculated. Reproducibility was summarized using coefficients of variation (CV) and intraclass correlation coefficients calculated from variance component models.
RESULTS
In healthy eyes, the CV of intravisit and intervisit global vessel density measures ranged from 1.8% to 3.2% in ONH scans and 2.5% to 9.0% in macular scans. In glaucoma eyes, the CV of intravisit and intervisit global vessel density measures ranged from 2.3% to 4.1% in ONH scans and 3.2% to 7.9% in macular scans. CVs were lower for global than sectorial measures. Global OCT-A ONH intraclass correlation measurements for the retinal nerve fiber layer in healthy eyes were lower (range: 0.65 to 0.85) than in glaucoma eyes (range: 0.89 to 0.94). Scan size did not make large differences in measurement CVs.
CONCLUSIONS
Reproducibility of OCT-A ONH and macula vessel density measurements is good. Moreover, glaucoma patients have sparser vessel density with poorer reproducibility than healthy subjects.
More by Patricia I C Manalastas
View full profile →Optical Coherence Tomography Angiography Macular Vascular Density Measurements and the Central 10-2 Visual Field in Glaucoma.
The Association Between Macula and ONH Optical Coherence Tomography Angiography (OCT-A) Vessel Densities in Glaucoma, Glaucoma Suspect, and Healthy Eyes.
Diagnostic Ability of Optical Coherence Tomography Angiography Macula Vessel Density for the Diagnosis of Glaucoma Using Difference Scan Sizes.
Top Research in Optic Nerve & Disc
Browse all →Efficacy of a Deep Learning System for Detecting Glaucomatous Optic Neuropathy Based on Color Fundus Photographs.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Inflammation in Glaucoma: From the back to the front of the eye, and beyond.
Discussion
Comments and discussion will appear here in a future update.