Long-term follow-up after implantation of a telemetric intraocular pressure sensor in patients with glaucoma: a safety report.
Summary
Telemetric IOP sensors showed good functionality and tolerability during long-term follow-up. Non-invasive self-tonometry with a telemetric IOP sensor can provide useful additional data for future monitoring of patients with glaucoma.
Abstract
IMPORTANCE
To investigate the long-term safety of a novel intraocular telemetric pressure sensor.
BACKGROUND
Acquisition of accurate intraocular pressure (IOP) data is vital for sufficient medical care of glaucoma patients. Non-invasive self-tonometry with a telemetric IOP sensor can provide important information regarding the individual IOP profile.
DESIGN
Retrospective analysis of long-term follow-up data assessed during outpatient visits in a university hospital.
PARTICIPANTS
Six patients with open-angle glaucoma were included. Unfortunately, one patient passed away shortly after completion of the original 1-year study.
METHODS
Within the scope of a prospective 1-year pilot clinical trial, a telemetric IOP sensor was inserted into the ciliary sulcus after intracapsular lens implantation during planned cataract surgery. Patients were regularly examined as outpatients even beyond the duration of the 1-year study. Data concerning sensor functionality, safety parameters and home self-tonometry were assessed.
MAIN OUTCOME MEASURES
Long-term sensor functionality and safety.
RESULTS
Sensor measurements were always successful in every patient. Additionally, home self-tonometry was conducted without any problems by every patient. The average follow-up period was 37.5 months (21-50 months). During this period, the average number of IOP measurements performed per patient was 1273 (223-2884 measurements). No severe adverse events were reported. A varying degree of pupillary distortion was observed after 6-12 months in every patient; this remained unchanged thereafter with only one exception.
CONCLUSIONS AND RELEVANCE
Telemetric IOP sensors showed good functionality and tolerability during long-term follow-up. Non-invasive self-tonometry with a telemetric IOP sensor can provide useful additional data for future monitoring of patients with glaucoma.
Keywords
More by Antonis Koutsonas
View full profile →Blue-yellow and standard pattern visual evoked potentials in phakic and pseudophakic glaucoma patients and controls.
Implantation of a novel telemetric intraocular pressure sensor in patients with glaucoma (ARGOS study): 1-year results.
Top Research in IOP & Medical Therapy
Browse all →The Complications of Myopia: A Review and Meta-Analysis.
Inflammation in Glaucoma: From the back to the front of the eye, and beyond.
Treatment Outcomes in the Primary Tube Versus Trabeculectomy Study after 1 Year of Follow-up.
Discussion
Comments and discussion will appear here in a future update.