A Case Report of Complete Blockage of a Baerveldt Glaucoma Implant Following Insertion of a 3-0 Supramid Suture.
Summary
We hypothesize that the blockage occurred at the junction between the device tube and plate and that the ballooning phenomenon observed was due to a defect in the tube wall.
Abstract
PURPOSE
The aim of this study was to present a case of a Baerveldt glaucoma implant lumen being completely occluded with a 3-0 Supramid stent suture.
PATIENT AND METHODS
The patient underwent Baerveldt glaucoma implant surgery with placement of an intraluminal 3-0 Supramid stent suture that acts to restrict flow across the device and reduce the risk of postoperative hypotony. Following suturing of the implant to the sclera, the device was flow tested. No flow was observed through the device tube and a significant ballooning of the tube diameter occurred with increased pressure on the device. The device was explanted from the eye and replaced with a different implant without further postoperative complication. The explanted device was assessed using custom microfluidic equipment in an in vitro environment.
RESULTS
This phenomenon occurred despite using several different batches of the 3-0 Supramid stent suture and the device had to be removed and replaced with another device without complication. In vitro microfluidic assessment of the device demonstrated no flow across the device tube despite over 150 mm Hg of pressure being exerted on the device.
CONCLUSIONS
We hypothesize that the blockage occurred at the junction between the device tube and plate and that the ballooning phenomenon observed was due to a defect in the tube wall. This case highlights the importance of flow testing all glaucoma drainage devices before insertion given the variation in manufacturing conditions to avoid the risk of intraoperative complications.
More by Richard M H Lee
View full profile →The Implications of an Ab Interno Versus Ab Externo Surgical Approach on Outflow Resistance of a Subconjunctival Drainage Device for Intraocular Pressure Control.
Genotype-Phenotype Associations of IL6 and PRG4 With Conjunctival Fibrosis After Glaucoma Surgery.
Author Response: Ab Interno versus Ab Externo Surgical Approach on Outflow Resistance of a Subconjunctival Drainage Device.
Discussion
Comments and discussion will appear here in a future update.