Anterior Chamber XEN Gel Stent Movements: The Impact on Corneal Endothelial Cell Density.
Summary
2% of the annual endothelial loss observed by Tan and colleagues in patients with a Baerveldt tube (4. 54% annual decrease).
Abstract
INTRODUCTION
Although there are several examples of glaucoma drainage devices' dislocations in the literature, the movement of XEN gel stents is scarcely documented, and the impact of such displacements on the corneal endothelium remains unknown. In light of the recent global market withdrawal of the CyPass microstent over concerns with regard to endothelial cell loss, we present a case of XEN gel stent dislocation into the anterior chamber, with localized corneal damage.
CASE REPORT
We describe the case of a 64-year-old female patient who underwent bilateral XEN gel stent implantation for normal-tension glaucoma. While right eye surgery was uneventful, intraocular pressure increased in the left eye, and examination revealed a flat filtration bleb with no microstent visible through the conjunctiva. A 3 mm segment of the XEN stent was visible within the anterior chamber, with corneal contact during eye movement and blinking. Endothelial cell density was measured twice 1 month apart before the XEN gel stent was replaced.
DISCUSSION
First, our observations confirm that XEN gel stents can dislocate and advance within the anterior chamber. We speculate that this displacement was caused by regular eye rubbing after surgery. Second, this report illustrates that the position of a drainage device within the anterior chamber is dynamic and significantly influenced by eye movements and blinking. Finally, occasional corneal contact of a XEN gel stent can cause a reduction in endothelial cell density of 2.1% in a month, which represents 46.2% of the annual endothelial loss observed by Tan and colleagues in patients with a Baerveldt tube (4.54% annual decrease).
More by Kevin Gillmann
View full profile →Combined and stand-alone XEN 45 gel stent implantation: 3-year outcomes and success predictors.
XEN Gel Stent in Pseudoexfoliative Glaucoma: 2-Year Results of a Prospective Evaluation.
Efficacy of Needling Revision After XEN Gel Stent Implantation: A Prospective Study.
Top Research in Glaucoma Surgery
Browse all →Treatment Outcomes in the Primary Tube Versus Trabeculectomy Study after 1 Year of Follow-up.
Efficacy, Safety, and Risk Factors for Failure of Standalone Ab Interno Gelatin Microstent Implantation versus Standalone Trabeculectomy.
Prospective, Randomized, Controlled Pivotal Trial of an Ab Interno Implanted Trabecular Micro-Bypass in Primary Open-Angle Glaucoma and Cataract: Two-Year Results.
Discussion
Comments and discussion will appear here in a future update.