Measurement Floors and Dynamic Ranges of OCT and OCT Angiography in Glaucoma.
Sasan Moghimi, Christopher Bowd, Linda M Zangwill, Rafaella C Penteado, Kyle Hasenstab, Huiyuan Hou, Elham Ghahari, Patricia Isabel C Manalastas, James Proudfoot, Robert N Weinreb
Summary
In late-stage glaucoma, particularly when VF MD is worse than -14 dB, OCTA-measured pfVD is a promising tool for monitoring progression because it does not have a detectable measurement floor.
Abstract
PURPOSE
To determine if OCT angiography (OCTA)-derived vessel density measurements can extend the available dynamic range for detecting glaucoma compared with spectral-domain (SD) OCT-derived thickness measurements.
DESIGN
Observational, cross-sectional study.
PARTICIPANTS
A total of 509 eyes from 38 healthy participants, 63 glaucoma suspects, and 193 glaucoma patients enrolled in the Diagnostic Innovations in Glaucoma Study.
METHODS
Relative vessel density and tissue thickness measurement floors of perifoveal vessel density (pfVD), circumpapillary capillary density (cpCD), circumpapillary retinal nerve fiber (cpRNFL) thickness, ganglion cell complex (GCC) thickness, and visual field (VF) mean deviation (MD) were investigated and compared with a previously reported linear change point model (CPM) and locally weighted scatterplot smoothing curves.
MAIN OUTCOME MEASURES
Estimated vessel density and tissue thickness measurement floors and corresponding dynamic ranges.
RESULTS
Visual field MD ranged from -30.1 to 2.8 decibels (dB). No measurement floor was found for pfVD, which continued to decrease constantly until very advanced disease. A true floor (i.e., slope of approximately 0 after observed CPM change point) was detected for cpRNFL thickness only. The post-CPM estimated floors were 49.5±2.6 μm for cpRNFL thickness, 70.7±1.0 μm for GCC thickness, and 31.2±1.1% for cpCD. Perifoveal vessel density reached the post-CPM estimated floor later in the disease (VF MD, -25.8±3.8 dB) than cpCD (VF MD, -19.3±2.4 dB), cpRNFL thickness (VF MD, -17.5±3.3 dB), and GCC thickness (VF MD, -13.9±1.8 dB; P < 0.001). The number of available measurement steps from normal values to the CPM estimated floor was greatest for cpRNFL thickness (8.9), followed by GCC thickness (7.4), cpCD (4.5), and pfVD (3.8).
CONCLUSIONS
In late-stage glaucoma, particularly when VF MD is worse than -14 dB, OCTA-measured pfVD is a promising tool for monitoring progression because it does not have a detectable measurement floor. However, the number of steps within the dynamic range of a parameter also needs to be considered. Although thickness parameters reached the floor earlier than OCTA-measured pfVD, there are more such steps with thickness than OCTA parameters.
More by Sasan Moghimi
View full profile →Development and Validation of a Deep Learning System to Detect Glaucomatous Optic Neuropathy Using Fundus Photographs.
Macular and Optic Nerve Head Vessel Density and Progressive Retinal Nerve Fiber Layer Loss in Glaucoma.
Optical Coherence Tomography Angiography Macular Vascular Density Measurements and the Central 10-2 Visual Field in Glaucoma.
Top Research in Visual Field
Browse all →Optical coherence tomography angiography: A comprehensive review of current methods and clinical applications.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Improving our understanding, and detection, of glaucomatous damage: An approach based upon optical coherence tomography (OCT).
Discussion
Comments and discussion will appear here in a future update.