Intraocular Degradation of XEN45 Gel Stent 3 Years After its Implantation.
Randolf A Widder, Pia Kühnrich, Matthias Hild, Corinna Rennings, Alexander Szumniak, Gernot F Rössler
Summary
Therefore, gonioscopy should be performed carefully to detect any changes in the stent.
Abstract
INTRODUCTION
We report a case of XEN45 Gel Stent removal 3 years after its implantation. The stent had degraded with a closed lumen, so a new stent was implanted.
CASE REPORT
A 63-year-old woman was scheduled for XEN45 Gel Stent implantation. Three years after the primary surgery and 2 years after a surgical revision, second revision surgery was required. The conjunctiva was opened, and the implanted XEN Gel Stent was stripped of the adhesive scar tissues. No aqueous humor flow was seen through the stent, and therefore, the stent was removed. The intrascleral and intracameral parts had degraded. The surface of the stent was irregular and cobbled. A part of the lumen in the middle of the stent had closed.
DISCUSSION
We detected changes in the stent after its removal during the surgical revision with conjunctival opening, which is our standard approach for failed XEN Gel Stents. Most surgeons advise needling of the stent in cases of failure. However, in cases of needling, changes in the stent might not have been detected. Thus, the number of patients with similar changes as this report might be higher than the publications suggest. No unique characteristics distinguished this patient from those who had a functional XEN Gel Stent. The stent, in this case, might have been of inferior quality as an exception, or all stents might degrade over time. Therefore, gonioscopy should be performed carefully to detect any changes in the stent. In cases of suspected stent degradation, we recommend a revision surgery with conjunctival opening and removal of the stent in cases of degradation.
More by Randolf A Widder
View full profile →The XEN45 Gel Stent as a minimally invasive procedure in glaucoma surgery: success rates, risk profile, and rates of re-surgery after 261 surgeries.
Predictability of ab-interno trabeculectomy success in the subsequent eye: A contralateral eye comparison study.
Primary Versus Posttrabeculectomy XEN45 Gel Stent Implantation: Comparison of Success Rates and Intraocular Pressure-lowering Potential in Pseudophakic Eyes.
Top Research in Glaucoma Surgery
Browse all →Treatment Outcomes in the Primary Tube Versus Trabeculectomy Study after 1 Year of Follow-up.
Efficacy, Safety, and Risk Factors for Failure of Standalone Ab Interno Gelatin Microstent Implantation versus Standalone Trabeculectomy.
Prospective, Randomized, Controlled Pivotal Trial of an Ab Interno Implanted Trabecular Micro-Bypass in Primary Open-Angle Glaucoma and Cataract: Two-Year Results.
Discussion
Comments and discussion will appear here in a future update.