Validation of a Head-mounted Virtual Reality Visual Field Screening Device.
Lukas Mees, Swati Upadhyaya, Pavan Kumar, Sandal Kotawala, Shankar Haran, Shruthi Rajasekar, David S Friedman, Rengaraj Venkatesh
Summary
While the CFA did not reliably identify deficits that matched the HFA, it was moderately effective at identifying glaucoma subjects.
Abstract
UNLABELLED
PRéCIS:: The C3 fields analyzer (CFA) is a moderately reliable perimeter preferred by patients to standard perimetry. While it does not approximate the gold standard, it was sensitive and specific for clinically defined glaucoma (area under the receiving operator characteristic curve=0.77 to 0.86).
PURPOSE
Testing the visual field is a vital sign for diagnosing and managing glaucoma. The current gold standard, the Humphrey visual field analyzer (HFA), is large, expensive and can be uncomfortable for some patients. The current study investigated the CFA, a virtual reality head-mounted visual field testing device, as a possible subjective field test for glaucoma screening and eventually glaucoma monitoring.
PATIENTS AND METHODS
The CFA presented stimuli in the same 54 positions as the HFA 24-2 SITA Standard test using a suprathreshold algorithm approximating an 18 dB deficit. A total of 157 patients (both controls and glaucoma patients) at the Aravind Eye Hospital, Pondicherry, India, were tested with both devices.
RESULTS
The number of stimuli missed on the CFA correlated with HFA mean deviation (r=0.62, P<0.001), and with pattern standard deviation (r=0.36, P<0.001). The area under the receiving operator characteristic curve was 0.77±0.06 for mild glaucoma (HFA mean deviation ≥-6 dB) and 0.86±0.04 for moderate-advanced glaucoma (HFA mean deviation <-6 dB). Patients with an 18 dB or worse deficit at a point in the visual field on the HFA failed to see the CFA stimulus at the same position 38% of the time.
CONCLUSIONS
While the CFA did not reliably identify deficits that matched the HFA, it was moderately effective at identifying glaucoma subjects. Further refinements to the device will be required to improve point by point testing performance and screening performance.
Top Research in Visual Field
Browse all →Optical coherence tomography angiography: A comprehensive review of current methods and clinical applications.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Improving our understanding, and detection, of glaucomatous damage: An approach based upon optical coherence tomography (OCT).
Discussion
Comments and discussion will appear here in a future update.