Global Search

Search articles, concepts, and chapters

J Cataract Refract SurgMay 198633 citations

Pathologic findings of an explanted silicone intraocular lens.

Newman D A, McIntyre D J, Apple D J, Deacon J, Popham J K, Isenberg R


AI Summary

An explanted silicone IOL causing glaucoma and blurred vision showed grooves, flash, and opalescence. Its removal resolved symptoms, highlighting IOL quality's impact on patient outcomes.

Abstract

We present a clinicopathological analysis of an explanted STAAR model B silicone intraocular lens (IOL) that was sent to the Center for Intraocular Lens Research for evaluation. Extracapsular cataract extraction (ECCE), using phacoemulsification and insertion of the silicone IOL, had been accomplished through a 3-mm scleral tunnel incision. The IOL was folded and inserted into the ciliary sulcus. Complications, including blurred vision, movement of the IOL within the eye, and glaucoma, eventually led to IOL exchange at three months post-ECCE. After removal of the silicone IOL, a modified J-loop IOL was placed in the intact capsular bag, with subsequent resolution of the increased intraocular pressure and a 20/20 + 3 visual acuity. Pathologic examination of the explanted silicone lens revealed grooves indented into the optic, extensive molding flash, and opalescence of the optic.


MeSH Terms

AgedGlaucomaHumansLenses, IntraocularMaleMicroscopy, Electron, ScanningPostoperative ComplicationsProsthesis DesignProsthesis FailureSiliconesVision DisordersVisual Acuity

Is this article assigned to the wrong chapter(s)? Let us know.