Results of First In Vivo Trial of an Acoustic Self-Tonometer.
Jan Osmers, Oskar Hoppe, Alicja Strzalkowska, Piotr Strzalkowski, Ágnes Patzkó, Stefan Arnold, Michael Sorg, Andreas Fischer
Summary
The usability and success rate of producing valid measurement values with the device during self-measurements by test subjects was nearly 92%.
Abstract
PURPOSE
Glaucoma is the world's most common cause of irreversible blindness, which makes early diagnosis, with the goal of preserving vision, essential. The current medical intervention is to reduce intraocular pressure (IOP) to slow down progression of the disease. The main goal of this study was to test a novel handheld acoustic self-tonometer on humans.
METHODS
A sound pressure pulse generated by a loudspeaker causes the eye to vibrate. A pressure chamber is placed on the human orbit to form a coupled system comprised of the patient's eye, the enclosed air, and the loudspeaker. A displacement sensor in front of the loudspeaker membrane allows the dynamic behavior of the entire system to be detected.
RESULTS
For this clinical trial series, a prototype of the acoustic self-tonometer principle was applied. The resulting membrane oscillation data showed sensitivity of patient IOP, but direct allocation of the measured damping and frequency to the IOP was not significant. For this reason, an artificial neural network was used to find relationships among the subjects' biometric eye parameters in combination with the self-tonometer data for the IOP reference. An expanded measurement uncertainty (= 2) equal to 6.53 mm Hg was determined for the self-tonometer in a Bland-Altman analysis using Goldmann applanation tonometer reference measurements.
CONCLUSIONS
The usability and success rate of producing valid measurement values with the device during self-measurements by test subjects was nearly 92%. The cross-sensitivities observed require compensation in a possible redesign phase to reduce the measurement uncertainty by at least 25% to the maximum of 5 mm Hg required to seek medical device approval.
TRANSLATIONAL RELEVANCE
Building on successful laboratory experiments with pig eyes, this article reports the results of testing the acoustic tonometer on humans.
Keywords
Top Research in IOP & Medical Therapy
Browse all →The Complications of Myopia: A Review and Meta-Analysis.
Inflammation in Glaucoma: From the back to the front of the eye, and beyond.
Treatment Outcomes in the Primary Tube Versus Trabeculectomy Study after 1 Year of Follow-up.
In the Knowledge Library
Discussion
Comments and discussion will appear here in a future update.