Agreement between rebound (Icare ic200) and applanation tonometry (Perkins) in patients with primary congenital glaucoma.
Laura Morales-Fernandez, Pilar Pérez-García, Federico Saenz-Frances, Mercedes Molero-Senosiain, Sofia Garcia-Saenz, Mendez Carmen Dora, Bueso Enrique Santos, Julian Garcia-Feijoo, Jose M Martinez-de-la-Casa
Summary
Despite the good overall agreement between both tonometers, caution should be taken in high values of IOP, considering the interchangeability of its readings as systematic and proportional differences appear to exist between both methods.
Abstract
PURPOSE
To examine agreement between intraocular pressure (IOP) measurements made using the rebound tonometer Icare ic200 (RT200) and the Perkins handheld applanation tonometer (PAT) in patients with primary congenital glaucoma (PCG). The impacts of several covariables on measurements using the two devices were also assessed.
MATERIALS AND METHODS
Intraocular pressure measurements were made in a single session in 86 eyes of 86 patients with PCG (46 under anaesthesia, 40 in the office). The order was RT200 then PAT. The variables age, central corneal thickness (CCT), corneal state and anaesthesia were recorded in each patient. Data were compared by determining interclass correlation coefficients (ICC) for each tonometer and representing the differences detected as Bland-Altman plots. Effects of covariables were assessed through univariate and multivariate regression.
RESULTS
Mean IOP difference between tonometers (RT200 minus PAT) was 1.26 mmHg (95%: 0.22-2.31). Absolute agreement (ICC) was 0.73 (95%
CI
0.62-0.82). Lower and upper limits of agreement (95%) were -8.06 mmHg (95%
CI
-9.87 to -6.25) and 10.59 mmHg (95%
CI
8.77-12.40), respectively. The tonometers showed systematic differences (a = -4.63 mmHg; 95%
CI
-9.11 to -1.44) and proportional differences; for each mmHg increase in PAT-IOP, the RT200 reading increased by 1.28 mmHg (b = 1.28; 95%
CI
1.12-1.53). None of the variables tested as predictors were able to explain differences between the tonometers.
CONCLUSIONS
Despite the good overall agreement between both tonometers, caution should be taken in high values of IOP, considering the interchangeability of its readings as systematic and proportional differences appear to exist between both methods.
Keywords
More by Laura Morales-Fernandez
View full profile →Tear and aqueous humour cytokine profile in primary open-angle glaucoma.
Changes to Corneal Topography and Biometrics After PRESERFLO Microshunt Surgery for Glaucoma.
Corneal Properties in Primary Open-angle Glaucoma Assessed Through Scheimpflug Corneal Topography and Densitometry.
Top Research in IOP & Medical Therapy
Browse all →The Complications of Myopia: A Review and Meta-Analysis.
Inflammation in Glaucoma: From the back to the front of the eye, and beyond.
Treatment Outcomes in the Primary Tube Versus Trabeculectomy Study after 1 Year of Follow-up.
In the Knowledge Library
Discussion
Comments and discussion will appear here in a future update.