Complications of Cosmetic Artificial Iris Implantation and Post Explantation Outcomes.
Summary
Cosmetic iris implantation was associated with serious complications at the time of presentation, and adverse sequelae persisted for years after explantation.
Abstract
PURPOSE
To report complications of cosmetic artificial iris implantation and explantation outcomes.
DESIGN
Retrospective case series.
METHODS
Medical records of 12 patients (24 eyes) who presented to us after being implanted with cosmetic artificial irises elsewhere were reviewed. Data collected included baseline demographics, presenting symptoms, examination findings, and management outcomes.
RESULTS
Eight eyes had NewColorIris implants and 16 had BrightOcular implants. The mean interval from cosmetic iris implantation to presentation was 61.7 ± 60.0 months. The mean follow-up after explantation was 35.5 ± 38.1 months. Complications at presentation included iris abnormalities (11 eyes, 45.8%), elevated intraocular pressure (8 eyes, 33.3%), corneal edema (6 eyes, 25%), intraocular inflammation (5 eyes, 20.8%), and cataract (4 eyes, 16.7%). Surgical interventions included cosmetic iris removal (19 eyes, 79.2%), cataract extraction (7 eyes, 29.2%), corneal transplantation (7 eyes, 29.2%), and glaucoma surgery (4 eyes, 16.7%). Complications at the last follow-up examination included native iris defects (11 eyes, 45.8%), persistent glaucoma (7 eyes, 29.2%), cataract (5 eyes, 20.8%), corneal edema (4 eyes, 16.7%), and intraocular inflammation (2 eyes, 8.3%). The mean logarithm of the minimum angle of resolution was 0.56 ± 0.47 at presentation and 0.78 ± 0.88 at the last examination (P = .30). The mean intraocular pressure was 22.7 ± 15.8 mm Hg at presentation and 13.4 ± 6.99 mm Hg at the last examination (P = .02).
CONCLUSION
Cosmetic iris implantation was associated with serious complications at the time of presentation, and adverse sequelae persisted for years after explantation.
More by Reza Ghaffari
View full profile →Top Research in Glaucoma Surgery
Browse all →Treatment Outcomes in the Primary Tube Versus Trabeculectomy Study after 1 Year of Follow-up.
Efficacy, Safety, and Risk Factors for Failure of Standalone Ab Interno Gelatin Microstent Implantation versus Standalone Trabeculectomy.
Prospective, Randomized, Controlled Pivotal Trial of an Ab Interno Implanted Trabecular Micro-Bypass in Primary Open-Angle Glaucoma and Cataract: Two-Year Results.
Discussion
Comments and discussion will appear here in a future update.