Three-year effectiveness and safety of the XEN gel stent as a solo procedure or in combination with phacoemulsification in open-angle glaucoma: a multicentre study.
Herbert Reitsamer, Vanessa Vera, Simon Ruben, Leon Au, Jorge Vila-Arteaga, Miguel Teus, Markus Lenzhofer, Andrew Shirlaw, Zhanying Bai, Mini Balaram, Ingeborg Stalmans
Summary
The gel stent effectively lowered IOP and IOP-lowering medication count over 3 years, with a predictable and acceptable safety profile, when implanted via the traditional ab-interno technique.
Abstract
PURPOSE
To assess the 3-year effectiveness and safety of the XEN gel stent implanted ab interno in open-angle glaucoma (OAG).
METHODS
This study was a multicentre, retrospective chart review of consecutive patients with OAG who underwent ab-interno gel stent placement alone or combined with phacoemulsification between 1 January 2014 and 1 October 2015. Outcome measures included mean changes in intraocular pressure (IOP) and IOP-lowering medication count from medicated baseline at 1, 2, 3 (primary outcome) and 4 years (if available) postimplantation. Intraoperative complications, adverse events of special interest (AESIs) and secondary surgical interventions (SSIs) were recorded.
RESULTS
The safety and effectiveness populations included 212 eyes (primary and secondary) and 174 eyes (primary), respectively. Mean IOP and medication decreased from 20.7 mmHg and 2.5 at baseline (n = 163 primary/first implanted eyes) to 13.9 mmHg and 1.1 medications (n = 76) at 3 years postimplantation, respectively. Mean changes from baseline in IOP (-5.6, -6.2 and -6.6 mmHg) and IOP-lowering medication count (-1.8, -1.6 and -1.4) were statistically significant at 1, 2 and 3 years postimplantation, respectively. Results appeared comparable when implantation was performed with (n = 76) or without (n = 98) phacoemulsification. In primary eyes with 4-year IOP and medication count data (n = 27), mean IOP was 14.0 mmHg on 1.3 medications at 4 years postimplantation. Fifteen (7.1%) eyes had intraoperative complications, 31 (14.6%) experienced 46 postoperative AESIs, and 26 (12.3%) required SSI.
CONCLUSION
The gel stent effectively lowered IOP and IOP-lowering medication count over 3 years, with a predictable and acceptable safety profile, when implanted via the traditional ab-interno technique.
Keywords
More by Herbert Reitsamer
View full profile →Efficacy, Safety, and Risk Factors for Failure of Standalone Ab Interno Gelatin Microstent Implantation versus Standalone Trabeculectomy.
Four-year results of a minimally invasive transscleral glaucoma gel stent implantation in a prospective multi-centre study.
Longitudinal bleb morphology in anterior segment OCT after minimally invasive transscleral ab interno Glaucoma Gel Microstent implantation.
Top Research in Glaucoma Surgery
Browse all →Treatment Outcomes in the Primary Tube Versus Trabeculectomy Study after 1 Year of Follow-up.
Efficacy, Safety, and Risk Factors for Failure of Standalone Ab Interno Gelatin Microstent Implantation versus Standalone Trabeculectomy.
Prospective, Randomized, Controlled Pivotal Trial of an Ab Interno Implanted Trabecular Micro-Bypass in Primary Open-Angle Glaucoma and Cataract: Two-Year Results.
Discussion
Comments and discussion will appear here in a future update.