A Method for Reducing the Number of Presentations in Perimetric Test Procedures.
Summary
ARBON can be added to a perimetry test procedure to speed up the test while largely preserving the distribution of returned sensitivities, thus producing normative data similar to the data for the original, underlying perimetric…
Abstract
PURPOSE
To introduce a new method (ARBON) for decreasing the test time of psychophysical procedures and examine its application to perimetry.
METHODS
ARBON runs in parallel with an existing psychophysical procedure injecting occasional responses of seen or unseen into that procedure. Using computer simulation to mimic human responses during perimetry, we assess the performance of ARBON relative to an underlying test procedure and a version of that procedure truncated to be faster. Simulations used 610 normal eyes (age 20 to 80 years) and 163 glaucoma eyes (median mean deviation = -1.81 dB, 5th percentile = +2.14 dB, 95th percentile = -22.55 dB). Outcome measures were number of presentations and mean absolute error in threshold estimation. We also examined the probability distribution of measured thresholds.
RESULTS
ARBON and the Truncated procedure reduced presentations by 16% and 18%, respectively. Mean error was increased by 8% to 10% for the Truncated procedure but decreased by 5% to 7% for ARBON. The probability distributions of measured thresholds using ARBON overlapped with the Underlying procedure by over 80%, whereas the Truncated procedure overlapped by 50%.
CONCLUSIONS
ARBON offers a principled method for reducing test time. ARBON can be added to any existing psychophysical procedure without requiring any change to the logic or parameters controlling the procedure, resulting in distributions of measured thresholds similar to those of the underlying procedure.
TRANSLATIONAL RELEVANCE
ARBON can be added to a perimetry test procedure to speed up the test while largely preserving the distribution of returned sensitivities, thus producing normative data similar to the data for the original, underlying perimetric test.
More by Andrew Turpin
View full profile →A Comparison between the Compass Fundus Perimeter and the Humphrey Field Analyzer.
Improving Spatial Resolution and Test Times of Visual Field Testing Using ARREST.
Orientation of the Temporal Nerve Fiber Raphe in Healthy and in Glaucomatous Eyes.
Top Research in Visual Field
Browse all →Optical coherence tomography angiography: A comprehensive review of current methods and clinical applications.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Improving our understanding, and detection, of glaucomatous damage: An approach based upon optical coherence tomography (OCT).
Discussion
Comments and discussion will appear here in a future update.