Virtual Reality Oculokinetic Perimetry Test Reproducibility and Relationship to Conventional Perimetry and OCT.

Visual FieldOCT & Imaging

Summary

A new virtual reality perimetry test for glaucoma shows reproducible results comparable to conventional tests, offering a quicker way to assess visual function in clinic or at home.

Abstract

PURPOSE

Vivid Vision Perimetry (VVP; Vivid Vision, Inc) is a novel method for performing in-office and home-based visual field assessment using a virtual reality platform and oculokinetic perimetry. Here we examine the reproducibility of VVP Swift and compare results with conventional standard automated perimetry (SAP) and spectral-domain (SD) OCT.

DESIGN

Cross-sectional study.

PARTICIPANTS

Fourteen eyes of 7 patients with open-angle glaucoma (OAG) (average age, 64.6 years; 29% women) and 10 eyes of 5 patients with suspected glaucoma (average age, 61.8 years; 40% women) were enrolled.

METHODS

Patients with OAG and suspected glaucoma were enrolled prospectively and underwent 2 VVP Swift examinations. Results were compared with 1 conventional SAP examination (Humphrey Visual Field [HVF]; Zeiss) and 1 SD OCT examination.

MAIN OUTCOME MEASURES

Mean sensitivity (in decibels) obtained for each eye in 2 VVP Swift test sessions and a conventional SAP examination, thickness of the retinal nerve fiber layer (RNFL) and ganglion cell complex (GCC) for the SD OCT examination, and mean test durations of the VVP Swift and SAP examinations.

RESULTS

The mean test duration of VVP Swift in both eyes (8.5 minutes) was significantly shorter (< 0.001) than SAP (12.2 minutes). The average absolute difference of the mean sensitivity between the 2 VVP Swift sessions was found to be 0.73 dB (95% confidence interval [CI], 0.40-1.06). A statistically significant association was found between average mean sensitivity measurements from the VVP and mean deviation (MD) measurements obtained by the HVF with a Pearson correlation coefficient of 0.86 (95% CI, 0.70-0.94;< 0.001). Mean visual sensitivity measurements from the VVP Swift test were significantly associated with average RNFL thickness ( = 0.66; = 0.014) and GCC thickness ( = 0.63; = 0.02), whereas the correlation coefficients between HVF MD and RNFL and GCC were 0.86 (< 0.001) and 0.83 (< 0.001), respectively.

CONCLUSIONS

Our results demonstrated that the VVP Swift test can generate reproducible results and is comparable with conventional SAP. This suggests that the device can be used by clinicians to assess visual function in glaucoma.

Keywords

CI, confidence intervalGCC, ganglion cell complexGlaucomaHVF, Humphrey Visual FieldMD, mean deviationOAG, open-angle glaucomaOCTOculokinetic perimetryPerimetryRNFL, retinal nerve fiber layerSAPSAP, standard automated perimetrySD, spectral-domainTele-ophthalmologyVVP, Vivid Vision PerimetryVirtual realityVisual field