Evaluation of the Consistency of Glaucomatous Visual Field Defects Using a Clustered SITA-Faster Protocol.
Jeremy C K Tan, Jack Phu, Denise Go, Doan Nguyen, Katherine Masselos, Allan Bank, Michael Kalloniatis, Ashish Agar
Summary
Frontloading SFR tests can provide repeatable data for the evaluation of the consistency of pattern deviation defects in glaucoma, with no observable decline in performance from test fatigue.
Abstract
PURPOSE
Frontloading SITA-Faster (SFR) visual fields (2 tests per eye on the same visit) has been shown to provide repeatable perimetric data at minimal time cost. This study reports the outcomes of using frontloaded SFR in the evaluation of pointwise visual field (VF) defects in a cohort of patients with glaucoma when transitioned from SITA-Standard (SS).
DESIGN
Prospective, cross-sectional study.
PARTICIPANTS
A total of 144 eyes of 91 patients with confirmed or suspected glaucoma who had an SS test on a previous visit.
METHODS
Two SFR tests (T1, T2) per eye on the same visit.
MAIN OUTCOME MEASURES
Global sensitivity, reliability indices, and pointwise deviation map probability scores from the pattern deviation grid of each patient were compared across the 3 sequential tests to evaluate the consistency of VF defects.
RESULTS
The mean age was 68.6 years, and 79.2% of patients had a diagnosis of glaucoma. There was no significant difference in mean deviation (MD) across the 3 tests (-5.83 decibels [dB], -5.28 dB, and -5.71 dB in SS, SFR1, and SFR2, respectively, repeated-measures analysis of variance [ANOVA], P = 0.48). The frontloaded SFR tests provided repeatable VFs that confirmed existing pointwise data on the SS in 4661 (62.3%) locations, reversed an SS defect in 614 (8.2%) locations, and demonstrated a new repeatable defect in 406 (5.4%) locations of the pattern deviation grid. A new defect of at least 3 contiguous points was identified in 20.1% of eyes. The non-repeatable points on the 2 SFR tests displayed no significant difference in the distribution of defect/nondefect points based on test order or peripheral versus central locations. There was no significant difference in the rate of obtaining at least 1 reliable test result between SS and the frontloaded SFR T1 and T2 (P = 0.77). Test duration significantly decreased from SS to SFR1/2 (379 vs. 160 vs. 158 seconds, P < 0.0001).
CONCLUSIONS
Frontloading SFR tests can provide repeatable data for the evaluation of the consistency of pattern deviation defects in glaucoma, with no observable decline in performance from test fatigue. This is achieved at equivalent duration and reliability as a single SS test. Frontloading SFR may be helpful in increasing testing frequency/quantity to meet recommended guidelines for progression analysis. FINANCIAL DISCLOSURE(S): Proprietary or commercial disclosure may be found in the Footnotes and Disclosures at the end of this article.
Keywords
More by Jeremy C K Tan
View full profile →Clinical Registries in Ophthalmology.
Frontloading SITA-Faster Can Increase Frequency and Reliability of Visual Field Testing at Minimal Time Cost.
Quantification and Predictors of Visual Field Variability in Healthy, Glaucoma Suspect, and Glaucomatous Eyes Using SITA-Faster.
Top Research in Visual Field
Browse all →Optical coherence tomography angiography: A comprehensive review of current methods and clinical applications.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Improving our understanding, and detection, of glaucomatous damage: An approach based upon optical coherence tomography (OCT).
In the Knowledge Library
Discussion
Comments and discussion will appear here in a future update.