Clin Exp Ophthalmol
Clin Exp Ophthalmol2024Comparative Study

Matched case-control comparison of surgical success after XEN45 Gel Stent and PRESERFLO MicroShunt implantation in a Caucasian population.

Glaucoma SurgeryIOP & Medical Therapy

Summary

Both the Preserflo and XEN45 Gel Stent provide an effective and safe treatment option for advanced glaucoma and have a high potential to reduce intraocular pressure.

Abstract

BACKGROUND

The aim of this study was to compare the outcome of implantation of a XEN45 Gel Stent with the outcome of implantation of a Preserflo MicroShunt in a matched-pair analysis in eyes being naïve to filtering surgery.

METHODS

In this comparative, retrospective study, 50 eyes that had undergone XEN45 Gel Stent implantation were compared with 50 eyes after Preserflo implantation. Follow-up was at least 6 months, and surgical success was measured by criteria A (IOP 20%, no repeat surgery); criteria B (IOP 20%, no repeat surgery); and criteria C (IOP ≤15 mmHg, IOP reduction ≥40%, no repeat surgery).

RESULTS

After a follow-up period of 12 months, mean IOP had decreased from preoperative 25.2 ± 4.8 mmHg in the XEN group to 14.5 ± 4.0 (n = 35) and from 25.3 ± 6.8 mmHg to 11.9 ± 2.9 (n = 41) in the Preserflo group, respectively. The IOP at the last follow-up of the two groups differed significantly (p < 0.01). The probability of surgical success did not differ concerning Criteria A and B, but surgical success was significantly higher in the Preserflo group for Criteria C (60%, p < 0.01).

CONCLUSION

Both the Preserflo and XEN45 Gel Stent provide an effective and safe treatment option for advanced glaucoma and have a high potential to reduce intraocular pressure. Absolute IOP levels of <16 mmHg after 12 months were significantly more frequent in the Preserflo group.