Uveitis-Glaucoma-Hyphema Syndrome From Subtle Malposition of a Hydrus Microstent.
Summary
Herein, we discuss a case of UGH syndrome due to the subtle malposition of a Hydrus microstent that ultimately required surgical removal to manage.
Abstract
Uveitis-glaucoma-hyphema (UGH) syndrome is a rare condition characterized by intraocular inflammation, elevated intraocular pressure, and recurrent hyphema. Although UGH syndrome is typically caused by chafing between an intraocular lens and the iris or ciliary body, microinvasive glaucoma devices are becoming increasingly recognized as potential causes of recurrent intraocular inflammation. Herein, we discuss a case of UGH syndrome due to the subtle malposition of a Hydrus microstent that ultimately required surgical removal to manage. This represents the first case of Hydrus-related UGH syndrome to include high-quality anterior segment imaging, which led us to suspect Hydrus-iris contact and supported the decision to proceed with microstent explantation.
More by Ashley Polski
View full profile →Top Research in Glaucoma Surgery
Browse all →Treatment Outcomes in the Primary Tube Versus Trabeculectomy Study after 1 Year of Follow-up.
Efficacy, Safety, and Risk Factors for Failure of Standalone Ab Interno Gelatin Microstent Implantation versus Standalone Trabeculectomy.
Prospective, Randomized, Controlled Pivotal Trial of an Ab Interno Implanted Trabecular Micro-Bypass in Primary Open-Angle Glaucoma and Cataract: Two-Year Results.
In the Knowledge Library