Comparison of Glaucoma Detection Performance of Binocular Perimetry Screening Program Using imo Perimetry With Frequency Doubling Technology.
Euido Nishijima, Daisuke Hosaka, Shumpei Ogawa, Yoshinori Itoh, Takahiko Noro, Sachiyo Okude, Kei Sano, Keiji Yoshikawa, Masayuki Tatemichi, Tadashi Nakano
Summary
The ISP demonstrates non-inferiority in detecting glaucoma and has a shorter testing time compared with FDT. These findings provide evidence for applied further studies on large-scale population-based glaucoma screening.
Abstract
PURPOSE
We aimed to preliminarily compare the glaucoma detection accuracy of a head-mounted binocular visual perimeter "imo" screening program (ISP) with that of frequency doubling technology (FDT).
METHODS
This multicenter, diagnostic accuracy study based on prospectively collected data included 76 non-glaucoma (including pre-perimetric glaucoma) eyes and 92 glaucomatous eyes from patients visiting two hospitals. Patients underwent ISP and FDT (C-20-1 screening program) on the same day. Diagnostic efficacy was evaluated using receiver operating characteristic curves and areas under the curve (AUCs). In addition, we compared the ISP and FDT testing times.
RESULTS
AUC values for ISP versus FDT were as follows: (1) mild-stage glaucoma (mean deviation [MD] > -6 dB), 0.82 (95% confidence interval [CI], 0.75-0.88) versus 0.76 (95% CI, 0.68-0.83); moderate-stage glaucoma (-6 dB ≥ MD ≥ -12 dB), 0.98 (95% CI, 0.95-1.00) versus 0.96 (95% CI, 0.93-1.00); and advanced-stage glaucoma (-12 dB > MD), 1.00 (95% CI, 1.00-1.00) versus 0.99 (95% CI, 0.98-1.00). In addition, mild-stage glaucoma was classified into two stages (MD > -3 D) and (-3 D ≥ MD > -6 D). AUC values were 0.81 (95% CI, 0.73-0.88) versus 0.76 (95% CI, 0.68-0.84) for MD > -3 D and 0.86 (95% CI, 0.77-0.94) versus 0.73 (95% CI, 0.61-0.86) for -3 D ≥ MD > -6 D. The testing time for the ISP was significantly shorter than that of FDT for all glaucoma stages (P < 0.001).
CONCLUSIONS
The ISP demonstrates non-inferiority in detecting glaucoma and has a shorter testing time compared with FDT. These findings provide evidence for applied further studies on large-scale population-based glaucoma screening.
TRANSLATIONAL RELEVANCE
Our study provides a non-inferior and quicker glaucoma screening than existing tools.
More by Euido Nishijima
View full profile →Comparative Analysis of 24-2C, 24-2, and 10-2 Visual Field Tests for Detecting Mild-Stage Glaucoma With Central Visual field Defects.
A Novel Visual Field Screening Program for Glaucoma With a Head-Mounted Perimeter.
Correlation of Binocular Perimetry Screening Using imo: A Portable Head Mounted Perimeter With 10-2 Standard Automated Perimetry for Early Glaucoma With Central Visual Field Defects.
Top Research in Visual Field
Browse all →Optical coherence tomography angiography: A comprehensive review of current methods and clinical applications.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Improving our understanding, and detection, of glaucomatous damage: An approach based upon optical coherence tomography (OCT).
In the Knowledge Library
Discussion
Comments and discussion will appear here in a future update.