Diagnostic Performance of the Offline Medios Artificial Intelligence for Glaucoma Detection in a Rural Tele-Ophthalmology Setting.
Swati Upadhyaya, Divya Parthasarathy Rao, Srinivasan Kavitha, Ganeshrao Shonraj Ballae, Kalpa Negiloni, Shreya Bhandary, Florian M Savoy, Rengaraj Venkatesh
Summary
Medios offline AI exhibited promising sensitivity and specificity in detecting referable glaucoma from remote vision centers in southern India when compared with teleophthalmologists.
Abstract
PURPOSE
This study assesses the diagnostic efficacy of offline Medios Artificial Intelligence (AI) glaucoma software in a primary eye care setting, using nonmydriatic fundus images from Remidio's Fundus-on-Phone (FOP NM-10). Artificial intelligence results were compared with tele-ophthalmologists' diagnoses and with a glaucoma specialist's assessment for those participants referred to a tertiary eye care hospital.
DESIGN
Prospective cross-sectional study
PARTICIPANTS
Three hundred three participants from 6 satellite vision centers of a tertiary eye hospital.
METHODS
At the vision center, participants underwent comprehensive eye evaluations, including clinical history, visual acuity measurement, slit lamp examination, intraocular pressure measurement, and fundus photography using the FOP NM-10 camera. Medios AI-Glaucoma software analyzed 42-degree disc-centric fundus images, categorizing them as normal, glaucoma, or suspect. Tele-ophthalmologists who were glaucoma fellows with a minimum of 3 years of ophthalmology and 1 year of glaucoma fellowship training, masked to artificial intelligence (AI) results, remotely diagnosed subjects based on the history and disc appearance. All participants labeled as disc suspects or glaucoma by AI or tele-ophthalmologists underwent further comprehensive glaucoma evaluation at the base hospital, including clinical examination, Humphrey visual field analysis, and OCT. Artificial intelligence and tele-ophthalmologist diagnoses were then compared with a glaucoma specialist's diagnosis.
MAIN OUTCOME MEASURES
Sensitivity and specificity of Medios AI.
RESULTS
Out of 303 participants, 299 with at least one eye of sufficient image quality were included in the study. The remaining 4 participants did not have sufficient image quality in both eyes. Medios AI identified 39 participants (13%) with referable glaucoma. The AI exhibited a sensitivity of 0.91 (95% confidence interval [CI]: 0.71-0.99) and specificity of 0.93 (95%
CI
0.89-0.96) in detecting referable glaucoma (definite perimetric glaucoma) when compared to tele-ophthalmologist. The agreement between AI and the glaucoma specialist was 80.3%, surpassing the 55.3% agreement between the tele-ophthalmologist and the glaucoma specialist amongst those participants who were referred to the base hospital. Both AI and the tele-ophthalmologist relied on fundus photos for diagnoses, whereas the glaucoma specialist's assessments at the base hospital were aided by additional tools such as Humphrey visual field analysis and OCT. Furthermore, AI had fewer false positive referrals (2 out of 10) compared to the tele-ophthalmologist (9 out of 10).
CONCLUSIONS
Medios offline AI exhibited promising sensitivity and specificity in detecting referable glaucoma from remote vision centers in southern India when compared with teleophthalmologists. It also demonstrated better agreement with glaucoma specialist's diagnosis for referable glaucoma participants. FINANCIAL DISCLOSURE(S): Proprietary or commercial disclosure may be found in the Footnotes and Disclosures at the end of this article.
Keywords
More by Swati Upadhyaya
View full profile →Validation of a Head-mounted Virtual Reality Visual Field Screening Device.
Assessment of Reasons for Presentation in New Primary Glaucoma Patients and Identification of Risk Factors for Late Presentation.
Letter to the Editor: The Impacts of Face Mask Use on Standard Automated Perimetry Results in Glaucoma Patients.
Top Research in Diagnosis & Screening
Browse all →Efficacy of a Deep Learning System for Detecting Glaucomatous Optic Neuropathy Based on Color Fundus Photographs.
Dry eye disease and oxidative stress.
Central Corneal Thickness in the Ocular Hypertension Treatment Study (OHTS).
In the Knowledge Library
Discussion
Comments and discussion will appear here in a future update.