Negative Pressure Application by the Ocular Pressure Adjusting Pump to Lower Intraocular Pressure in Normal-Tension Glaucoma: HERCULES Study.
Tanner J Ferguson, Thomas W Samuelson, Leon Herndon, Daniel C Terveen, Jason Bacharach, Jacob W Brubaker, John P Berdahl, Nathan M Radcliffe
Summary
The ocular pressure adjusting pump safely and effectively lowers both daytime and nocturnal IOP in patients with NTG.
Abstract
PURPOSE
To evaluate the safety and intraocular pressure (IOP)-lowering efficacy of the ocular pressure adjusting pump in subjects with normal-tension glaucoma (NTG).
DESIGN
Prospective, multicenter, masked, randomized, fellow-eye controlled trial. SUBJECTS, PARTICIPANTS,
AND/OR CONTROLS
Subjects with NTG with an IOP ≥12 mm Hg and ≤21 mm Hg were enrolled. One eye of each subject was randomized to receive negative pressure application; the fellow eye served as a control.
METHODS
Subjects wore the device overnight for 1 year and the applied negative pressure was programmed by subtracting a reference IOP of 6 mm Hg from the baseline IOP.
MAIN OUTCOME MEASURES
The primary effectiveness endpoint was the proportion of eyes achieving an IOP reduction ≥20% at Week 52 during the day. The secondary endpoint was the proportion of eyes achieving a nocturnal IOP reduction ≥20% at Week 52. Exploratory endpoints included mean IOP reduction in clinic and in the sleep lab.
RESULTS
A total of 186 eyes were randomized across 11 sites. 120 eyes successfully completed all visits across 52 weeks without protocol deviations. At Week 52, 88.3% (n = 53) of study eyes vs 1.7% (n = 1) of control eyes met the primary endpoint. For the secondary endpoint, 96.7% (n = 58) of study eyes vs 5.0% (n = 3) met the endpoint. For exploratory IOP analysis, the mean nocturnal IOP reduction at Week 52 was 8.0 mm Hg (39.1%) from a baseline of 20.4 ± 2.5 mm Hg to 12.4 ± 2.7 mm Hg. There were no serious adverse events. The most commonly reported adverse events were lid (11.8% study, 1.1% control) and periorbital edema (12.9%, 1.1%).
CONCLUSIONS
The ocular pressure adjusting pump safely and effectively lowers both daytime and nocturnal IOP in patients with NTG.
More by Tanner J Ferguson
View full profile →Trabecular microbypass stent implantation with cataract extraction in pseudoexfoliation glaucoma.
Evaluation of a Trabecular Micro-Bypass Stent in Pseudophakic Patients With Open-Angle Glaucoma.
Trabecular microbypass stent implantation in pseudophakic eyes with open-angle glaucoma: Long-term results.
Top Research in IOP & Medical Therapy
Browse all →The Complications of Myopia: A Review and Meta-Analysis.
Inflammation in Glaucoma: From the back to the front of the eye, and beyond.
Treatment Outcomes in the Primary Tube Versus Trabeculectomy Study after 1 Year of Follow-up.
In the Knowledge Library
Discussion
Comments and discussion will appear here in a future update.