One-year follow-up of posterior chamber toric phakic intraocular lens implantation for moderate to high myopic astigmatism.
Kamiya Kazutaka, Shimizu Kimiya, Aizawa Daisuke, Igarashi Akihito, Komatsu Mari, Nakamura Akio
AI Summary
Toric ICLs effectively and safely corrected moderate to high myopic astigmatism over one year, showing excellent visual outcomes and stability, making them a viable surgical option.
Abstract
Objective
To assess the 1-year clinical outcomes of toric Visian Implantable Collamer Lens (ICL; STAAR Surgical, Nidau, Switzerland) implantation for moderate to high myopic astigmatism.
Design
Prospective, observational case series.
Participants
Fifty-six eyes of 32 consecutive patients, with spherical equivalent errors of -4.00 to -17.25 diopters (D) and cylindrical errors of -0.75 to -4.00 D, who underwent toric ICL implantation.
Methods
Before and 1 week and 1, 3, 6, and 12 months after surgery, the safety, efficacy, predictability, stability, and adverse events of the surgery were assessed in eyes undergoing toric ICL implantation. Ocular higher-order aberrations (HOAs) and contrast sensitivity (CS) function also were evaluated before and 1 year after surgery.
Main outcome measures
Uncorrected visual acuity (UCVA), best spectacle-corrected visual acuity (BSCVA), safety index, efficacy index, predictability, stability, adverse events, HOAs, and CS function.
Results
The logarithm of the minimum angle of resolution (logMAR) UCVA and logMAR BSCVA were -0.11 (corresponding to Snellen equivalent 20/16) ± 0.12 and -0.19 (corresponding to 20/12.5) ± 0.08 1 year after surgery, respectively. The safety and efficacy indices were 1.17 ± 0.21 and 1.00 ± 0.29. At 1 year, 91% and 100% of the eyes were within 0.5 and 1.0 D, respectively, of the targeted correction. Manifest refraction changes of -0.07 ± 0.27 D occurred from 1 week to 1 year. For a 4-mm pupil, fourth-order aberrations were changed, not significantly, from 0.05 ± 0.02 μm before surgery to 0.06 ± 0.03 μm after surgery (P = 0.38, Wilcoxon signed-rank test). Similarly, for a 6-mm pupil, fourth-order aberrations were not significantly changed, merely from 0.20 ± 0.08 μm before surgery to 0.23 ± 0.11 μm after surgery (P = 0.15). The area under the log CS function was significantly increased from 1.41 ± 0.15 before surgery to 1.50 ± 0.13 after surgery (P < 0.001). No vision-threatening complications occurred during the observation period.
Conclusions
In the authors' experience, the toric ICL performed well in correcting moderate to high myopic astigmatism during a 1-year observation period, suggesting its viability as a surgical option for the treatment of such eyes.
MeSH Terms
Shields Classification
Related Articles5
Postoperative complications of intrascleral intraocular lens fixation: A systematic review and meta-analysis.
Systematic ReviewBoston Type II Keratoprosthesis, 7 Years of Experience and Outcomes From a Spanish Tertiary Eye Care Hospital.
Case SeriesLong-term efficacy and safety profiles after posterior chamber phakic intraocular lens implantation in eyes with more than 10 years of follow-up.
Retrospective StudyLONG-TERM OUTCOME OF TRANSSCLERAL FOUR-POINT FIXATION OF AKREOS INTRAOCULAR LENS WITH CLOSED CONTINUOUS-LOOP SUTURE.
Case SeriesU.S. military implantable collamer lens surgical outcomes: 11-year retrospective review.
Observational StudyIs this article assigned to the wrong chapter(s)? Let us know.