Macular and retinal nerve fiber layer thickness measurements in normal eyes with the Stratus OCT, the Cirrus HD-OCT, and the Topcon 3D OCT-1000.
Huang Jingjing, Liu Xing, Wu Ziqiang, Guo Xinxing, Xu Hongzhi, Dustin Laurie, Sadda Srinivas
AI Summary
This study compared three OCT devices, finding retinal thickness measurements are not interchangeable and reproducibility varies, meaning clinicians must account for device differences when monitoring glaucoma.
Abstract
Purpose
To compare retinal nerve fiber layer (RNFL) and macular thickness measurements using time-domain optical coherence tomography (TD OCT) and 2 Fourier-domain OCTs (FD OCT).
Methods
Sixty eyes of 60 normal participants underwent eye examination followed by OCT evaluation with the time-domain Stratus OCT (Carl Zeiss Meditec, Inc. Dublin, CA) and 2 FD OCTs-the Cirrus HD-OCT (Carl Zeiss Meditec, Inc. Dublin, CA) and the 3D OCT-1000 (Topcon Corporation, Tokyo, Japan). Each patient was scanned 3 times on each machine by the same physician examiner. RNFL and macular thickness measurements were compared across the 3 devices using repeated measures analysis of variance (ANOVA) and multiple comparison t tests with Bonferroni adjustments. The coefficients of variation (CV) and the intraclass correlation coefficients were calculated to assess reproducibility. Bland-Altman plots were constructed to assess the level of agreement between every 2 OCTs.
Results
All 3 machines revealed similar patterns of regional differences in RNFL and macular thicknesses. Macular thickness was consistently greatest on the Cirrus and lowest on the Stratus device. However, RNFL measurements were less consistent. The overall average RNFL thickness was 4 μm greater on the Stratus than on the 3D OCT-1000, which was in turn 5 μm greater than on the Cirrus device. Bland-Altman plots showed poor-to-moderate agreement in terms of macular and RNFL thickness measurements between each 2 of the OCT devices. Both FD OCTs provided more reproducible readings on macular measurements than the Stratus, as evidenced by the lower CV and higher intraclass correlation coefficients. However, RNFL measurements on the 2 FD OCTs were not necessarily more reproducible than those from the Stratus OCT.
Conclusion
Thickness measurements are not interchangeable among different OCT devices because of the poor-to-moderate interdevice measurement agreement. FD instruments yield more reproducible macular but not RNFL thickness measurements.
MeSH Terms
Shields Classification
Related Articles5
Evaluation of visual function and OCT parameters in ethambutol-induced optic neuropathy: a longitudinal study.
Cohort StudyPrimary Visual Pathway Changes in Individuals With Chronic Mild Traumatic Brain Injury.
Case-Control StudyLongitudinal Changes of Retinal Nerve Fiber Layer and Ganglion Cell-Inner Plexiform Layer in Highly Myopic Glaucoma: A 3-Year Cohort Study.
Cohort StudyEvaluating glaucoma in myopic eyes: Challenges and opportunities.
ReviewRetinal Perfusion and Injury in Sepsis and after Major Surgery.
Observational StudyIs this article assigned to the wrong chapter(s)? Let us know.