Agreement among three types of spectral-domain optical coherent tomography instruments in measuring parapapillary retinal nerve fibre layer thickness.
Kanamori Akiyasu, Nakamura Makoto, Tomioka Mari, Kawaka Yuki, Yamada Yuko, Negi Akira
AI Summary
This study found that three OCT devices measure RNFL thickness differently, with Cirrus showing smaller values. Despite correlations, they are not interchangeable due to significant measurement discrepancies.
Abstract
Backgrounds/aims: To evaluate the agreement of parapapillary retinal nerve fibre layer (RNFL) thickness among three spectral-domain optical coherence tomography (OCT) instruments.
Methods
Two hundred and three glaucomatous eyes and 88 normal eyes were imaged by Cirrus, RTVue and 3D OCT. The average and the four quadrant RNFL thicknesses were evaluated. Agreement among RNFL measurements was evaluated using Bland-Altman analysis and linear regression analysis. The percentage of each quadrant in the average RNFL thickness value was compared among the three instruments.
Results
Cirrus showed significantly smaller thickness values than RTVue (difference=8.8 μm, p<0.0001) and 3D OCT (difference=8.1 μm, p<0.0001). Although RNFL measurements among the instruments were highly correlated, the Bland-Altman analysis revealed proportional biases for most of the pair-wise agreements. Additionally, 3D OCT showed strong proportional biases with RTVue and 3D OCT. RTVue had a smaller occupied proportion of nasal quadrants (30.2%) and a larger proportion of inferior quadrants (32.4%) compared with Cirrus and 3D OCT.
Conclusions
RNFL measurements among the instruments were well correlated but had different values for thickness. The measurement circle of RTVue might be more superior-temporally located compared with the other instruments. Differences in the measurement protocols might be affected by the disagreements. These instruments should not be used interchangeably.
MeSH Terms
Shields Classification
Related Articles5
Longitudinal Changes of Retinal Nerve Fiber Layer and Ganglion Cell-Inner Plexiform Layer in Highly Myopic Glaucoma: A 3-Year Cohort Study.
Cohort StudyIndependent Evaluation of RETFound Foundation Model's Performance on Optic Nerve Analysis Using Fundus Photography.
Observational StudyProximal Location of Optic Disc Hemorrhage and Glaucoma Progression.
Cohort StudyPrediction of Visual Field Progression with Baseline and Longitudinal Structural Measurements Using Deep Learning.
Observational StudyNumber of macula optical coherence tomography scans needed to detect glaucoma progression.
Cohort StudyIs this article assigned to the wrong chapter(s)? Let us know.