J Glaucoma
J Glaucoma2015Research Support, Non-U.S. Gov't

Combining Frequency Doubling Technology Perimetry and Scanning Laser Polarimetry for Glaucoma Detection.

Diagnosis & ScreeningGlaucoma Surgery

Summary

The multivariable model including GDx-TSNIT, NAP-FDT, and the interaction GDx-TSNIT×NAP-FDT provides the best glaucoma prediction compared with all other multivariable and univariable models.

Abstract

PURPOSE

To determine the ability of frequency doubling technology (FDT) and scanning laser polarimetry with variable corneal compensation (GDx-VCC) to detect glaucoma when used individually and in combination.

METHODS

One hundred ten normal and 114 glaucomatous subjects were tested with FDT C-20-5 screening protocol and the GDx-VCC. The discriminating ability was tested for each device individually and for both devices combined using GDx-NFI, GDx-TSNIT, number of missed points of FDT, and normal or abnormal FDT. Measures of discrimination included sensitivity, specificity, area under the curve (AUC), Akaike's information criterion (AIC), and prediction confidence interval lengths.

RESULTS

For detecting glaucoma regardless of severity, the multivariable model resulting from the combination of GDx-TSNIT, number of abnormal points on FDT (NAP-FDT), and the interaction GDx-TSNIT×NAP-FDT (AIC: 88.28,

AUC

0.959, sensitivity: 94.6%, specificity: 89.5%) outperformed the best single-variable model provided by GDx-NFI (AIC: 120.88,

AUC

0.914, sensitivity: 87.8%, specificity: 84.2%). The multivariable model combining GDx-TSNIT, NAP-FDT, and interaction GDx-TSNIT×NAP-FDT consistently provided better discriminating abilities for detecting early, moderate, and severe glaucoma than the best single-variable models.

CONCLUSIONS

The multivariable model including GDx-TSNIT, NAP-FDT, and the interaction GDx-TSNIT×NAP-FDT provides the best glaucoma prediction compared with all other multivariable and univariable models. Combining the FDT C-20-5 screening protocol and GDx-VCC improves glaucoma detection compared with using GDx or FDT alone.

Discussion