J Glaucoma
J GlaucomaAugust 2015Observational Study

Reference Standard Test and the Diagnostic Ability of Spectral Domain Optical Coherence Tomography in Glaucoma.

OCT & ImagingDiagnosis & Screening

Summary

Diagnostic parameters of SDOCT that was significantly affected by reference standard was the specificity, which was greater with optic disc classification as the reference standard.

Abstract

PURPOSE

To evaluate the relationship between the reference standard used to diagnose glaucoma and the diagnostic ability of spectral domain optical coherence tomograph (SDOCT).

METHODS

In a cross-sectional study, 280 eyes of 175 consecutive subjects, referred to a tertiary eye care center for glaucoma evaluation, underwent optic disc photography, visual field (VF) examination, and SDOCT examination. The cohort was divided into glaucoma and control groups based on 3 reference standards for glaucoma diagnosis: first based on the optic disc classification (179 glaucoma and 101 control eyes), second on VF classification (glaucoma hemifield test outside normal limits and pattern SD with P-value of 0.05). However, specificities of SDOCT parameters were significantly greater (P<0.05) with optic disc classification as reference standard (74% to 88%) compared with VF classification as reference standard (57% to 74%).

CONCLUSIONS

Diagnostic parameters of SDOCT that was significantly affected by reference standard was the specificity, which was greater with optic disc classification as the reference standard. This has to be considered when comparing the diagnostic ability of SDOCT across studies.